Section 3
Setting the Installation Distance
■
AUTO mode
Enter the current distance "d" between the workpiece and the Sensor Head.
Enter the sensor distance value again when the relationship between the static electric
sensor and the displacement sensor changes.
■
FIX mode
Fix the static electricity sensor (ZJ-SD) and the laser sensor (ZX-LD__) or static
electricity sensor (ZJ-SD) and the ultrasonic sensor (ZX-UD__) to the exclusive
mounting bracket (ZJ-XBU1). Simply setting the FIX mode frees you from the need to
input the distance value.
Workpiece
Static electricity
sensor
ZJ-SD
Displacement
sensor
"d" input
(60 mm)
Amplifier automatically recognizes distance B
(A-d) between displacement sensor and static
electricity sensor by input of "d".
Change in distance with workpiece is sensed
based on this value, and electric potential
value is corrected.
A
B
Corrected electric potential value is displayed on amplifier according to distance correction
coefficient held by ZJ-S sensor.
Corrected electric potential value is displayed on amplifier according to distance correction
coefficient held by ZJ-S sensor.
Workpiece
Mounting bracket (ZJ-XBU1)
A
B
Displacement
sensor
Static electricity
sensor
ZJ-SD
Distance B is stored on amplifier beforehand by securing
onto exclusive mounting bracket.
Change in distance with workpiece is sensed based on
value of A-B, and electric potential value is corrected.
operation2.book Page 42 Wednesday, November 29, 2006 2:56 PM