
Auto Focus Multi Code Reader V430-series User’s Manual (Z392)
3-13
Communications
Response Timeout
Usage:
Only used when a response is required from the host. The reader can be
set to wait indefinitely by setting
Response Timeout
to zero.
Definition:
The time that the reader will wait before timing out if
ACK
,
NAK
, and
ETX
are enabled, and a host response is expected.
Serial Cmd:
<
K143,
response timeout
>
Default:
12
(in 1 ms increments)
Options:
0 to 255 (A zero (0) setting causes an indefinite wait.)
Summary of Contents for V430-F-series
Page 1: ...Auto Focus Multi Code Reader V430 F series User s Manual Z392 E1 02 ...
Page 9: ...CONTENTS 7 Auto Focus Multi Code Reader V430 series User s Manual Z392 ...
Page 125: ...5 28 Auto Focus Multi Code Reader V430 series User s Manual Z392 Good Decode s Needed ...
Page 279: ...9 14 Auto Focus Multi Code Reader V430 series User s Manual Z392 New Master Pin ...
Page 403: ...Omron Microscan Systems Inc 700 SW 39th St Suite 100 Renton WA 98057 USA ...