Manual Inflation Blood 

Pressure Monitor

Model S1

Instruction Manual

Thank you for purchasing the OMRON S1 Manual Inflation Blood Pressure 
Monitor.

The OMRON S1 is a compact manual inflation blood pressure monitor, 
operating on the oscillometric principle. It measures your blood pressure and 
pulse rate simply and quickly. 

Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR 
DOCTOR.

Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia 
or arteriosclerosis.
Please read this section carefully before using the unit.

Warning:   

• Indicates a potentially hazardous situation which, if not avoided, could result 

in death or serious injury.

(General Usage)

• Always consult your doctor. Self-diagnosis of measurement results and 

self-treatment are dangerous.

• People with severe blood flow problems, or blood disorders, should consult a 

doctor before using the unit, as cuff inflation can cause internal bleeding.

(Battery Usage)

• If battery fluid should get in your eyes, immediately rinse with plenty of clean 

water. Consult a doctor immediately.

Caution: 

• Indicates a potentially hazardous situation which, if not avoided, may result in 

minor or moderate injury to the user or patient or damage to the equipment or 
other property.

(General Usage)

• Do not leave the unit unattended with infants or persons who cannot express 

their consent.

• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result 

in incorrect measurement results.

• Make sure that the air tube is not wrapped around other parts of your body 

when taking measurements at night. This could result in injury when the air 
pressure in the air tube is increased.

• Do not leave the cuff wrapped on the arm if taking measurements during the 

night. This could result in injury.

• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, 

near the unit. This may result in incorrect operation of the unit.

• Do not operate the unit in a moving vehicle (car, airplane).

(Battery Usage)

• If battery fluid should get on your skin or clothing, immediately rinse with 

plenty of clean water.

• Use only two “AAA” alkaline or manganese batteries with this unit. Do not use 

other types of batteries.

• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace both batteries at 

the same time.

• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.

General Precautions

• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking, 

exercising or eating.

• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull the air tube with holding near the air jack.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic 

Compatibility (EMC)” in the Technical Data Section.

• Read and follow the “Correct Disposal of This Product” in the Technical Data 

Section when disposing of the device and any used accessories or optional 
parts.

Important Safety Information

Note: Subject to technical modification without prior notice. 

Important information regarding Electro Magnetic Compatibility (EMC)

With the increased number of electronic devices such as PC’s and mobile 
(cellular) telephones, medical devices in use may be susceptible to 
electromagnetic interference from other devices. Electromagnetic interference 
may result in incorrect operation of the medical device and create a potentially 
unsafe situation.
Medical devices should also not interfere with other devices.

In order to regulate the requirements for EMC (Electro Magnetic Compatibility) 
with the aim to prevent unsafe product situations, the EN60601-1-2:2007 
standard has been implemented. This standard defines the levels of immunity to 
electromagnetic interferences as well as maximum levels of electromagnetic 
emissions for medical devices.

This medical device manufactured by OMRON Healthcare conforms to this 
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate 

strong electrical or electromagnetic fields, near the medical device. This may 
result in incorrect operation of the unit and create a potentially unsafe situation. 
Recommendation is to keep a minimum distance of 7 m. Verify correct operation 
of the device in case the distance is shorter.

Further documentation in accordance with EN60601-1-2:2007 is available at 
OMRON Healthcare Europe at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.

This marking shown on the product or its literature, indicates that it 
should not be disposed of, with other household wastes at the end 
of its working life. To prevent possible harm to the environment or 
human health from uncontrolled waste disposal, please separate 
this from other types of wastes and recycle it responsibly to 
promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this 
product, or their local government office, for details of where and how they can 
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions 
of the purchase contract. This product should not be mixed with other commercial 
wastes for disposal. 

This product does not contain any hazardous substances.

Correct Disposal of This Product 
(Waste Electrical & Electronic Equipment)

Introduction

1.   Overview

Main unit: 

A. Display
B. M (Memory) button

C. I/O (Power switch) button
D. Air jacks for cuff and inflation bulb
E. Battery compartment

F. Arm cuff
 

(Medium cuff: arm 

 circumference 

22-32 

cm)

G. Air  Tube

H.  Air Release Button
I. Air 

Tube

J.  Air Inflation Bulb

Display: 

Arm cuff: 

O. Systolic blood pressure
P.  Diastolic blood pressure
Q. Pulse display
R.  Memory No. symbol
S. Deflation symbol

K. Heartbeat symbol

1.  Flashes during measurement
2.  If flashing after measurement 

complete indicates blood 
pressure out of recommended 
range

L. Reinflation symbol
M. Battery low symbol
N. Memory symbol

2.1   Installing/Replacing the Batteries

2.   Preparation

1.

2.

Remove the battery cover.

Insert two “AAA” batteries 
as indicated in the battery 
compartment and then 
replace the battery cover.

Notes:
• If the battery low symbol (     ) appears on the display, turn the unit off 

then replace both batteries at the same time.

• The measurement values continue to be stored in memory even after 

the batteries are replaced.

Disposal of used batteries should be carried out in accordance with the 
national regulations for the disposal of batteries.

The bottom edge of the cuff should be 1 to 2 cm above the elbow. 
Marker (arrow under tube) is centred on the middle of your inner arm. 
Close the fabric fastener FIRMLY.

3.   Using the Unit

1.

Insert the air tubes into the air jacks.

2.

Put your arm through the cuff loop.

3.

Position the arm correctly.

1.

2.

Pump the inflation bulb to inflate the arm cuff.

This criteria is for home blood pressure measurement.

Note:  If the memory is full, the monitor will delete the oldest readings.

Systolic Blood Pressure 

Above 135 mmHg 

Diastolic Blood Pressure 

Above 85 mmHg

3.1   Applying the Arm Cuff

The monitor automatically stores the result up to 14 sets.

3.3   Taking a Reading

6.   Technical Data

5.   Optional Parts

4.3   Maintenance

3.4   Using the Memory Function

Calibration and Service

To protect your unit from damage, please observe the following:

Notes:
• When you take a measurement on the right 

arm, air tube will be at the side of your 
elbow. Be careful not to rest your arm on 
the air tube.

Note:
If the deflation symbol does not disappear soon, press the air release 
button to release any air in the arm cuff.

To delete all the values stored in memory

• Do not subject the main unit, cuff and inflation bulb to 

extreme temperatures, humidity, moisture or direct 
sunlight.

• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations 

(for example, dropping the unit on the floor).

• Do not use volatile liquids to clean the main unit. The 

unit should be cleaned with a soft, dry cloth.

• Use a soft, moistened cloth and soap to clean the arm 

cuff.

• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean 

the arm cuff.

• The accuracy of this blood pressure monitor has been 

carefully tested and is designed for a long service life.

• It is generally recommended to have the unit inspected 

every two years to ensure correct functioning and 
accuracy. Please consult your authorised OMRON 
dealer or the OMRON Customer Service at the address 
given on the packaging or attached literature. 

• Do not carry out repairs of any kind yourself. If a defect 

occurs, consult your OMRON retail outlet or distributor as 
mentioned on the packaging.

Manual Inflation Blood Pressure Monitor
OMRON S1 (HEM-4030-RU)
LCD Digital Display
Oscillometric method
Pressure: 0 mmHg to 299 mmHg 
Pulse: 40 to 180/min.
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Manual by inflation bulb
Automatic pressure release valve
14 Measurements
2 “AAA” batteries 1.5V
Capacity of new manganese batteries is approx. 
1500 measurements

        = Type B           

Internally powered ME equipment

+10°C to +40°C / Maximum: 30 to 90% RH

-20°C to +60°C / Maximum: 10 to 95% RH / 
700-1060 hPa
Approximately 80g without batteries
Approximately 120g
Approximately 64 (w) mm x 35 (h) mm x 105 (I) mm
Approximately 146 mm x 446 mm 
(Medium cuff: arm circumference 22 to 32 cm)
Nylon and polyester
Main unit, cuff, inflation bulb, instruction manual, 
storage case, battery set, guarantee card, blood 
pressure pass

Inflation bulb: 

Storage case: 

Remove tight-fitting clothing or tight rolled up sleeve from your 
upper arm.
Do not place the cuff over thick clothes.

3.2   How to Sit Correctly

To take a measurement, you need to be 
relaxed and comfortably seated, under 
comfortable room temperature. No eating, 
smoking or exercising 30 minutes before 
taking a measurement.

• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your 

heart. 

1 to 2 cm

• The blood pressure can differ between the right arm and the left arm, 

and therefore also the measured blood pressure values can be 
different. Omron recommends to always use the same arm for 
measurement. If the values between the two arms differ substantially, 
please check with your physician which arm to use for your 
measurement.

Press the air release button to release any air in the arm 
cuff, then press the I/O button to turn the unit on.

3.

Measurement starts.
Measurement starts automatically after you stop inflating 
the arm cuff.

4.

Remove the arm cuff.

5.

Press the I/O button to turn off the monitor.
The monitor automatically stores the measurement in its 
memory.
It will automatically turn off after five minutes.

Notes:
• If the reinflation symbol (    ) appears, squeeze the inflation bulb to 

re-inflate the arm cuff.

• Do not inflate the arm cuff more than necessary.

Notes:
• To cancel a measurement, press the I/O button to turn off the unit and 

press the air release button to release the air in the arm cuff.

• Wait 2-3 minutes before taking another blood pressure measurement. 

Waiting between readings allows the arteries to return to the condition 
prior to taking the blood pressure measurement.

Note:  You cannot partially delete the stored readings.

When the memory symbol (    ) appears, first press the M button. 
Then while holding it down, press the I/O button simultaneously for 
about 2-3 seconds.

Note: Remain still while taking measurement.

Self-diagnosis of measured results and treatment are dangerous. 
Please follow the instructions of your doctor.

1)  Inflate the cuff until it is 30 to 40 mmHg 

above your expected systolic blood 
pressure value.

 

ex) If your expected blood pressure is 
around 140mmHg, inflate the arm cuff to 
between 170 and 180 mmHg. Inflate the 
cuff rapidly so that the pressure is reached in about five seconds.

2)  When the desired pressure has been achieved, release the inflation 

bulb.

START

DEFLATE

COMPLETE

Warning:

• If your systolic or diastolic pressure is outside 

the standard range, the heartbeat symbol will 
blink when the measurement result is 
displayed.

  Recent research suggests that the following 

values can be used as a guide to high blood 
pressure for measurements taken at home.

Important:

Previous Reading

4.2   Troubleshooting

Problem

Cause

Remedy

• Press the I/O button and repeat measurement.
• If the problem continues, try replacing the 

batteries with new ones.

If this still does not solve the problem, contact 
your OMRON retail outlet or distributor.

The reading is 
extremely low (or 
high).

Arm cuff pressure 
does not rise.

Cannot measure or 
readings are too low 
or too high.

The unit loses power 
during measurement.

Nothing happens 
when you press the 
buttons.

Other problems.

Arm cuff not applied 
correctly.

Movement or talking 
during measurement.

Clothing is interfering 
with the arm cuff.

The air tube is not 
securely connected 
into the main unit.

Air is leaking from the 
arm cuff.

The arm cuff has not 
been inflated 
sufficiently.

The air release button 
being pressed during 
inflation.

The batteries are 
empty.

The batteries are 
empty.

The batteries have 
been inserted 
incorrectly.

Apply the arm cuff 
correctly. 
Refer to section 3.1.

Remain still and do not 
talk during 
measurement.
Refer to section 3.3.

Remove any clothing 
interfering with the arm 
cuff.
Refer to section 3.2.

Make sure that the air 
tube is connected 
securely. 
Refer to section 3.1.

Replace the arm cuff 
with a new one. 
Refer to Chapter 5.

Inflate the cuff so that it 
is 30 to 40 mmHg above 
your previous 
measurement result.
Refer to section 3.3.

Be careful not to press 
the air release button 
during measurement.

Replace the batteries 
with new ones.
Refer to section 2.1.

Replace the batteries 
with new ones.
Refer to section 2.1.

Insert the batteries with 
the correct (+/-) polarity.
Refer to section 2.1.

4.   Troubleshooting and Maintenance

4.1   The Icons and Error Messages

Blink 

Lit 

The systolic blood pressure 
is above 135 mmHg or 
diastolic blood pressure is 
above 85 mmHg. 

The batteries are low.

The batteries are 
exhausted. 

Cuff is under inflated.

Movement during 
measurement.

Cuff over inflated.

Device error.

Pressure is too low.

Recent research suggests 
these values can be used 
as a guide to high blood 
pressure for measurements 
taken at home.

 

You should replace them 
with new ones ahead of 
time.
Refer to section 2.1.

You should replace them 
with new ones at once.
Refer to section 2.1.

Carefully read and repeat 
the steps listed under 
section 3.3.

Contact your OMRON retail 
outlet or distributor.
Refer to Chapter 6.

Cause

Remedy

Error Display

Press the inflation bulb to 
inflate the arm cuff until the 
reinflation symbol goes out. 
Or, deflate the arm cuff and 
repeat measurement after 
checking that the heartbeat 
symbol (     ) has been 
displayed.
Refer to section 3.3.

1.

Unplug the air tube from the air connector.

2.

Gently fold the air tube into the arm cuff.

4.4   Storage

Keep the unit in its storage case when not in use.
Do not keep the main unit being wound with the arm cuff.

Note:
Do not bend the air tube excessively. 

Note:
Should you require a small cuff, please ensure that the small cuff and 
small bulb are used together.
They may also be purchased as a combination set.

3.

Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct 

sunlight, dust or corrosive vapours.

• Locations exposed to vibrations, shocks or where it will be at an 

angle.

Place the arm cuff, inflation bulb and 
main unit in the storage case.

Medium Arm Cuff

Arm circumference 
22 - 32 cm

Large Arm Cuff

Arm circumference 
32 - 42 cm 

Small Arm Cuff

Arm circumference 
17 - 22 cm

CM-9515371-7

CL-9515370-9

CS2-9515373-3

Small cuff and bulb combination

Arm circumference 17 - 22 cm

Regular bulb

CSB2-9515372-5

4997965-1

Product Description
Model
Display
Measurement Method
Measurement Range

Accuracy

Inflation
Deflation
Memory
Power Source
Battery Life

Applied Part

Protection Against 
Electric Shock
Operating 
temperature/ Humidity
Storage temperature/ 
Humidity/Air pressure
Console Weight
Cuff Weight
Outer Dimensions
Cuff Dimensions

Cuff Material
Package Content

• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device 

Directive). 

• This blood pressure monitor is designed according to the European Standard 

EN1060, Non-invasive sphygmomanometers Part 1: General Requirements 
and Part 3: Supplementary requirements for electromechanical blood 
pressure measuring systems.

• This OMRON product is produced under the strict quality system of OMRON 

Healthcare Co. Ltd., Japan. The Core component for OMRON blood pressure 
monitors, which is the Pressure Sensor, is produced in Japan.

IM-HEM-4030-RU-02-10/2011

                     5325570-6C

Made in China

Manufacturer

OMRON HEALTHCARE Co., Ltd.

53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN

OMRON HEALTHCARE EUROPE B.V.

Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com

ComplectService Ltd.

13-14 building, 26 B. Tishinskiy per., 
Moscow 123557, RUSSIA
www.csmedica.ru

OMRON DALIAN CO., LTD.

Dalian, CHINA

EU-representative

Production Facility

12K2607

Exclusive
distributor in
Russia & importer

Reviews: