WARNING
•
Failure to properly clean the MICROAir components will result in the risk of bacterial
growth, serious illness from contamination, compromised medication delivery.
•
Clean the parts of this device before using this device for the first time; after this device
has not been used for an extended period of time and when parts are contaminated.
•
Clean the parts of this device after each use.
•
Allow the parts to dry after cleaning before reassembling and store in a clean location so
that this device will not become contaminated.
•
DO NOT leave the cleaning solution in the parts. Rinse the parts with clean tap water
after cleaning.
CAUTION
•
DO NOT push or poke at the mesh with items such as a cotton swab or pin.
•
Dry the parts by air-drying only. DO NOT use any equipment to dry the device parts.
•
To prevent damage to the device, DO NOT immerse the Main Unit in any liquid, DO NOT wash or
rinse any of the parts under strong running water, and DO NOT touch the mesh with your hand or any
object.
•
DO NOT use household bleach to clean the device or parts; doing so will cause rust and damage the
Mesh Cap.
•
To prevent damage to the device, DO NOT clean the Main Unit using abrasive cleaners or any type
of chemical.
•
NEVER clean with benzene or thinner.
Before Using the Device for the First Time
Before using the device for the first time, disassemble the parts and clean the mesh cap,
medication container, mouthpiece, mask, mask adapter and vibrating element and its
surrounding surface as shown below:
Mesh Cap
•
Wash the mesh cap in mild (neutral) detergent, then rinse off with distilled water.
•
Remove excess water and allow to air dry in a clean place. Some types of medications may remain
on the mesh.
•
Make sure to clean the medication from the mesh.
•
Do not wash under running water.
Model
NE-U100
Vibrating Mesh Nebulizer
How to Clean the Device
3609065-5A