IntelliSense

®

Blood Pressure Monitor

Model

HEM-907XL

Table of Contents

Be Sure to Read This Section

Exemptions.........................................2

Notes on Safety..................................3

Know Your Unit 

Features of the Product .....................7

Components of the Product ...............8

Features and Functions .....................9

Names of the Parts ........................9

Functions and setting 

while the unit is in use ..................11

Preparations before Measurement

How to Apply the Cuff ......................13

How to Use the Power Source.........15 

How to Use the AC Adapter

...........

16

Installation and Replacement 

of Battery Pack .............................17

How to Measure Blood Pressure

List of Measurement Modes ............18

SINGLE Mode ..............................19

AVG Mode ....................................21

MANU Mode .................................23

How to Check Pressure Accuracy..25

How to Clean the Unit after Use ........26

List of Error Codes ...........................27

Troubleshooting ................................28

Specifications

Caution.............................................29

Five Year Limited Warranty...............29

Specifications ...................................30

Thank you for purchasing the OMRON
IntelliSense

®

Blood Pressure Monitor.

Please thoroughly read this Instruction Manual
before using this monitor to ensure safe and
accurate use.

Please keep this manual near the monitor all
the time for future reference.

INSTRUCTION MANUAL

HEM-907-Z2_E_M03_090828.pdf

HEM-907-Z2_E_M.qxd  09.8.28  11:45 AM  Page 1

Summary of Contents for IntelliSense HEM-907XL

Page 1: ...ply the Cuff 13 How to Use the Power Source 15 How to Use the AC Adapter 16 Installation and Replacement of Battery Pack 17 How to Measure Blood Pressure List of Measurement Modes 18 SINGLE Mode 19 AVG Mode 21 MANU Mode 23 How to Check Pressure Accuracy 25 How to Clean the Unit after Use 26 List of Error Codes 27 Troubleshooting 28 Specifications Caution 29 Five Year Limited Warranty 29 Specificat...

Page 2: ...or modify this product 4 When Notes on Safety or Instructions for Use contained in this manual are not followed 5 When use and or operation of this device is effected by an act of God such as fire earthquake flood or other natural disasters 1 The product and contents of this Instruction Manual may be changed without prior notice 2 We have prepared the contents of this Instruction Manual thoroughly...

Page 3: ...ates something that is compulsory always follow Matters involving actual compulsory actions are indicated by text or pictures in or near The pictured icon refers to unplugging the power source plug Do not wrap the cuff over an arm to which intravenous injection or transfusion is being conducted or when otherwise contraindicated Do not connect the air tube or the cuff to other equipment which is co...

Page 4: ... not share an electric outlet with other unit or electric appliance After cleaning this unit dry it well before plugging the AC adapter in the electric outlet If this unit fails to perform as indicated discontinue use turn off the unit unplug the AC adapter from the electric outlet and contact OMRON s repair department Do not disassemble or modify this unit Do not use any cuff other than the model...

Page 5: ...ving the unit Read the instruction manual of the other devices to be used at the same time with this unit to understand and be aware of the interaction between the devices When using the unit Do not inflate the cuff without being wrapped over the arm Do not use a damaged cuff Be sure that patients do not touch the Buttons of this unit After using the unit Do not disinfect this unit by autoclave or...

Page 6: ... unit operates normally and safely before use Troubleshooting If device error 9 Er9 occurs take the following procedure promptly 1 Remove the cuff from the patient s arm 2 Turn off the power of the unit and unplug the AC adapter from the electric outlet 3 Display Out of use on this unit so that it cannot be used 4 Contact Omron Healthcare s Customer Service at 1 877 216 1336 ...

Page 7: ...d pressures and pulse rate are displayed Each measurement can also be shown individually The number of measurements waiting time before first measurement and the interval can be changed Auscultation Mode MANU Mode You can measure auscultatory blood pressure by using a stethoscope with automatic cuff inflation and deflation by this unit Because the cuff pressures during deflation are displayed digi...

Page 8: ...DE P SE T 100 AUTO 140 180 220 260 280 MODE AVG MANU CHECK SINGLE ON OFF STAR T STOP mmHg SYS mmHg DIA BPM PULSE AC AR T IND EX IND EX MA X MA X XL IntelliSense Blood Pressure Monitor ModelHEM 907XL Table of Contents Be Sure to Read This Section Exemptions 2 Notes on Safety 3 Know Your Unit Features of the Product 7 Components of the Product 8 Options 9 Features and Functions 10 Names of the Parts...

Page 9: ...eck Mode CHECK Mode Checks the accuracy of pressure display Displays only pressure 6 ON OFF power Button Turns on or off the unit 7 START Button Starts the measurement 8 DEFLATION deflation control Measurement Result Display Switch Button In the MANU Mode deflates the cuff by approximately 5 to 10 mmHg with each push during deflation In the AVG Mode switches the display of average values and the m...

Page 10: ...f irregular pulse waves The pulse wave having a difference of more than 25 from the average interval of pulse waves is determined irregular Number of irregular pulse waves is displayed up to three times Battery level Displays for the operable level Blinks for a low charged level Displays for the inoperable level Charging Displays when the battery pack is being charged External power source Display...

Page 11: ...setting If the cuff has not been inflated to the necessary level it may be re inflated automatically ON OFF AVG 1st 2nd 3rd DEFLATION START STOP HIDE P SET 100 AUTO 140 180 220 260 280 MODE AVG MANU CHECK SINGLE ON OFF AVG 1st 2nd 3rd DEFLATION START STOP HIDE P SET 100 AUTO 140 180 220 260 280 MODE AVG MANU CHECK SINGLE 2 Non display function Use to prohibit the display of measurement results How...

Page 12: ...surement interval for the AVG Mode Items to set Set value F1 Number of measurements 2 times or 3 times F2 Waiting time until the start of 1st measurement 0 sec 3 min 5 min or 10 min F3 Measurement interval 5 sec 30 sec 1 min or 2 min Note The bold letters represent the factory set values Procedure to change the set values 1 When the power is OFF push the ON OFF power Button for more than three sec...

Page 13: ... bladder set by rotating the luer connection as shown in the Figure on the right Do not wrap the cuff over an arm to which intravenous injection or transfusion is being conducted or when otherwise contraindicated Do not connect the air tube or the cuff to other equipment which is connected to intracorporeal organ Air embolisms may result The cuff of OMRON IntelliSense Blood Pressure Monitor HEM 90...

Page 14: ...ed 1 2 to 1 above the inner side of elbow joint If the INDEX is positioned outside the RANGE select the cuff suitable for the patient s arm circumference and wrap it again Wrap the cuff so that you can insert only one finger between the cuff and arm 6 Keep the level of the cuff at the same level as the heart during the measurement 3 Place the right or left hand of the patient with the palm of hand...

Page 15: ...for use with the device may damage the unit DANGER TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK CAREFULLY FOLLOW THESE INSTRUCTIONS For connection to a power supply not in the U S A use an attachment plug adapter of the proper configuration for the power outlet The power unit AC Adapter is intended to be correctly oriented in a vertical or floor mount position SAVE THESE INSTRUCTIONS ...

Page 16: ...o not touch the AC adapter with wet hands Be sure to use the AC adapter from the power supply of 110 VAC Do not install or store this unit where it may come in contact with water or liquid medication This is a Class II device with double isolation Earth pin is not for protective purposes 1 2 How to use the AC adapter Connect the AC adapter to the DC jack of the main unit 1 and the electric outlet ...

Page 17: ...quently replace it Charging time At approximately five seconds after connecting the AC adapter the unit will start battery charging automatically While the battery is being charged the icon turns on The battery can be completely charged in approximately twelve hours Battery low When the icon starts to blink twenty to thirty measurements remain on the battery However if the Battery Low Mark starts ...

Page 18: ...INGLE Mode Refer to Page 19 To measure two times or three times automatically and consecutively AVG Mode Refer to Page 21 To measure by using a stetho scope MANU Mode Refer to Page 23 Confirm readings with the stethoscope when an irregular pulse wave is displayed or when the measured value is questionable or erratic Do not use a cellular phone near this unit Do not use this unit in a vehicle How t...

Page 19: ...et the MODE Selector to SINGLE 3 Set the P SET inflation level Knob to AUTO or to the target pressure value 4 Measure the patient s arm size and wrap appropriate cuff over the patient s arm Refer to Pages 13 and 14 HOW TO MEASURE BLOOD PRESSURE IN SINGLE MODE ...

Page 20: ...ile the battery pack is in use the monitor will turn off automatically after five minutes of inactivity and the display measurement results will disappear Automatic Power Off 7 Push the ON OFF power Button to turn off the power If the monitor determines that the pressure value is not correct an error display appears Er1 to 9 In this case refer to Page 27 and start the measurement again HOW TO MEAS...

Page 21: ...s 2 Waiting time until the 1st measurement 0 sec Interval 1 min To change these factory set values refer to Page 12 3 Set the P SET inflation level setting Knob to AUTO or the target pressure value 4 Measure the patient s arm size and wrap appropriate cuff over the patient s arm Refer to Pages 13 and 14 HOW TO MEASURE BLOOD PRESSURE IN AVERAGE MODE ...

Page 22: ...er If the monitor determines that the pressure value is not correct an error display appears Er1 to 9 In this case refer to Page 27 and start the measurement again For setting the number of measurements the waiting time until the 1st measurement and the interval refer to Page 12 If you want to stop measurement push the STOP Button The unit will rapidly deflate If an error occurs during measurement...

Page 23: ...t the P SET inflation level setting Knob to AUTO or the target pressure value 4 Measure the patient s arm size and wrap appropriate cuff over the patient s arm Refer to pages 13 and 14 2 Set the MODE Selector to MANU HOW TO MEASURE BLOOD PRESSURE IN MANUAL MODE ...

Page 24: ... push the DEFLATION deflation control Measurement Results Display Switch Button Each time the Button is pushed cuff is deflated rapidly in increments of 5 to 10 mmHg 9 Push the ON OFF power Button to turn off the power 7 Take the readings 8 Push the STOP Button to remove air inside the cuff The unit does not automatically deflate in the MANU Mode If the monitor determines that the pressure value i...

Page 25: ...e s Cusomer Service at 1 877 216 1336 How to check 1 Connect the manometer inflation bulb cuff and the monitor with the T tube as shown in the figure on the right 2 Tightly wrap the cuff over a sturdy cylinder 3 Release the valve of inflation bulb to remove the air inside the cuff completely 4 Push the ON OFF power Button to turn on the monitor 5 Set the MODE Selector to CHECK 6 Close the valve of...

Page 26: ...s unit with gasoline paint thinner or high concentration alcohol Do not disinfect this unit by autoclave or gas steril ization EtO gluteraldehyde or high concentration ozone When cleaning this unit unplug the AC adapter from the electric outlet After cleaning this unit dry it well before plugging the AC adapter in the electric outlet HOW TO CLEAN THE UNIT AFTER USE HOW TO CLEAN THE UNIT AFTER USE ...

Page 27: ...stay within the range of 30 to 199 beats min Device error Main unit malfunction How to correct Confirm that the air tube connecting the cuff and the main unit is connected securely Confirm that the air flow in the air tube connecting the cuff and the main unit isn t being restricted Confirm that the cuff is wrapped correctly refer to pages 13 and 14 Check bladder for leaks and if necessary replace...

Page 28: ... the patient not to move during measurement and measure again Set the P SET to 30 to 40 mmHg higher than estimated systolic pressure of the patient then measure After checking the patient with the stethoscope refer to the list of error codes Refer to Page 27 Ask the patient not to move during measurement and measure again Set the P SET to 30 to 40 mmHg higher than estimated systolic pressure of th...

Page 29: ...could void the user s authority to operate this product NOTE POTENTIAL FOR RADIO TELEVISION INTERFERENCE for U S A only This product has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference in a residential installation The product generates uses and ...

Page 30: ...F to 104 F 10 to 40 C 30 to 85 RH IPX 0 Rating Weight of Main Unit Approx 32 oz 910 g External Dimensions Approx 5 1 2 W x 8 H x 5 1 6 D 139 W x 203 H x 131 D mm Accessories Cuff bladder set Extra Large Cuff bladder set Large Cuff bladder set Medium Cuff bladder set Small AC adapter Battery pack air tube 1 3 m 51 3 16 Instruction Manual Options Air tube 1 3 m 51 3 16 Stand exclusive for this unit ...

Page 31: ...31 NOTE ...

Page 32: ... Toll Free 1 877 216 1336 Made in Japan Manufactured by OMRON HEALTHCARE Distributed by OMRON HEALTHCARE INC 1200 Lakeside Drive Bannockburn Illinois 60015 www omronhealthcare com Copyright 2009 OMRON HEALTHCARE INC 1628519 0E ...