Solar Charging

Manual Inflation Blood Pressure Monitor

Model HEM-SOLAR

• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
• 

РУКОВОДСТВО

 

ПО

 

ЭКСПЛУАТАЦИИ

EN

FR

DE

IT

ES

NL

RU

IM-HEM-4500-SOLE-01-03/09

5322504-1A

AR

Rev.10 2008/11/28

HEM-4500-SOLE_main.book  Page 1  Tuesday, February 3, 2009  1:30 PM

Summary of Contents for HEM-SOLAR

Page 1: ... Pressure Monitor Model HEM SOLAR Instruction Manual Mode d emploi Gebrauchsanweisung Manuale di istruzioni Manual de instrucciones Gebruiksaanwijzing РУКОВОДСТВО ПО ЭКСПЛУАТАЦИИ EN FR DE IT ES NL RU IM HEM 4500 SOLE 01 03 09 5322504 1A AR ...

Page 2: ...structions 3 Using the Unit 17 3 1 How to Sit Correctly When Taking a Measurement 17 3 2 Applying the Arm Cuff 19 3 3 Taking a Reading 22 4 Quick Reference Guide 26 Care and maintenance 5 Handling Errors and Problems 27 5 1 Error Messages 27 5 2 Troubleshooting 29 6 Maintenance and Storage 31 7 Optional Parts 33 8 Technical Data 34 9 Some Useful Information about Blood Pressure 37 ...

Page 3: ...ng on the oscillometric principle It measures your blood pressure and pulse rate simply and quickly The OMRON HEM SOLAR also recharges the battery using Solar Power Allowing for use in various situations and locations Please read this instruction manual thoroughly before using the unit For specific information about your own blood pressure CONSULT YOUR DOCTOR Before using the unit ...

Page 4: ...eyes immediately rinse with plenty of clean water Consult a doctor immediately AC Adapter Usage Never plug in or unplug the power cord from the electric outlet with wet hands Caution Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property General Usage Do not leave the unit unattende...

Page 5: ...ned for this unit Please refer to Chapter 7 AC Adapter Usage Use only the original AC adapter designed for this unit Use of unsupported adapters may damage and or may be hazardous to the unit Plug the AC adapter into the appropriate voltage outlet Do not use the AC adapter if the unit or the power cord is damaged Turn off the power and unplug the power cord immediately General Precautions Do not o...

Page 6: ... Important information regarding Electro Magnetic Compatibility EMC in the Technical Data Section Read and follow the Correct Disposal of This Product in the Technical Data Section when disposing of the device and any used accessories or optional parts ...

Page 7: ...B I O button Power switch C Air connector cap for cuff D Air connector for cuff E Air connector for inflation bulb F Air connector cap for inflation bulb G AC Adapter jack H AC Adapter jack cap I Battery compartment J Solar panel A B I J D C E F G H ...

Page 8: ...8 1 Overview Arm Cuff Inflation Bulb K Arm cuff Medium cuff arm circumference 22 32 cm L Air Tube M Air Plug N Air Release Button O Air Tube P Air Inflation Bulb Q Air Plug L K M O P N Q ...

Page 9: ...tor W Heartbeat symbol 1 Flashes during measurement 2 If flashing after measurement completed indicates blood pressure out of recommended range X Pulse display Y Battery low symbol Note If your systolic or diastolic pressure is outside the standard range above 135 85 mmHg the Heartbeat symbol will blink Please refer to Chapter 3 3 R W Y S T U X V ...

Page 10: ...10 1 Overview Package contents others Z Storage case AA AC adapter AB Rechargeable battery Instruction manual Guarantee card Blood pressure pass AA AB Z ...

Page 11: ...dapter then charge up the battery Connecting the Rechargeable Battery 1 Turn the main unit upside down 2 Slide the battery cover in the direction of the arrow while pressing the ribbed part of the cover 3 Connect the Rechargeable battery to the main unit Note When replacing the Rechargeable battery pull the cables and disconnect the Rechargeable battery from the main unit ...

Page 12: ...utlet 6 After the battery is fully charged remove the AC adapter plug from the monitor and close the AC adapter jack cap Notes The orange LED is lit while the battery is being charged by the AC adapter After the battery is fully charged the orange LED goes out Charge up the Rechargeable battery with the AC adapter before taking a measurement for the first time or in case the Rechargeable battery b...

Page 13: ...ashing then lights up when the Rechargeable battery is empty In case of using two AAA size batteries together with the Rechargeable battery the battery low symbol starts flashing then lights up when the AAA batteries are exhausted Use two disposable AAA size batteries to avoid the unit turning off after the Rechargeable battery becomes empty This devices operates on rechargeable battery two AAA si...

Page 14: ...C adapter jack cap Charge the battery inside with the direct rays of the sun near a window The battery won t be charged by room light Do not leave the unit outdoors and avoid getting the unit wet from rain If the Rechargeable battery life becomes extremely short it might be a temporary problem with the capacity Charge the battery with the AC adapter To keep the Rechargeable battery capacity longer...

Page 15: ...Note The capacity might change depending upon the weather situation Weather Lux To fully charge the battery 4 hours continuous chargewithasouthern exposure Clear more than 80000lux 35 hours continuous charge 40 measurements Fair 50000lux 90 hours continuous charge 20 measurements Cloudy 20000lux 600 hours continuous charge 5 measurements ...

Page 16: ...ks into place If the battery low symbol appears on the display replace both batteries at the same time When the battery low symbol starts to blink you will still be able to use the unit for a short while You should replace the batteries with new ones ahead of time When the symbol remains lit the batteries are exhausted You should replace the batteries with new ones at once Remove the batteries if ...

Page 17: ...ohol smoking or exercising for at least 30 minutes before taking a measurement Do not move or talk during measurement Correct Posture Remove tight fitting clothing from your upper arm and any thick clothing such as a sweater Do not place the cuff over thick clothes and do not roll up your sleeve if it is too tight Sit upright with your back straight Place your arm on a table so that the cuff will ...

Page 18: ...re values due to strain or the arm cuff being lower than the heart If the arm cuff is at a lower position than your heart use cushions etc to adjust the height of your arm Arched back leaning forwards Sitting cross legged Sitting on a sofa or at a low table so that you tend to lean forward ...

Page 19: ...efore also the measured blood pressure values can be different Omron recommends to always use the same arm for measurement If the values between the two arms differ substantially please check with your doctor which arm to use for your measurement Taking measurements on the left arm 1 Open two Air connector caps then insert the air plugs of the arm cuff and the inflation bulb into their respective ...

Page 20: ...p 3 Position the arm correctly 4 When the cuff is positioned correctly close the fabric fastener FIRMLY 1 The air tube should run down the inside of your forearm and be in line with your middle finger 2 The bottom of the cuff should be approximately 1 to 2 cm above your elbow 3 Apply the cuff to your upper arm so that the coloured marker blue arrow under tube is centered on the middle of your inne...

Page 21: ...ents on the right arm Apply the cuff so that the air tube is at the side of your elbow Be careful not to rest your arm on the air tube or otherwise restrict the flow of air to the cuff The cuff should be 1 to 2 cm above the elbow 1 to 2 cm ...

Page 22: ...lation symbol does not disappear soon press the air release button to release any air in the arm cuff 2 Pump the inflation bulb to inflate the arm cuff 1 Inflate the cuff until it is 30 to 40 mmHg above your expected systolic blood pressure value ex If your expected blood pressure is around 140mmHg inflate the arm cuff to between 170 and 180 mmHg Inflate the cuff rapidly so that the pressure is re...

Page 23: ...e display and the heartbeat symbol flashes Remain still and do not talk during measurement Note To cancel a measurement press the I O button to turn off the unit and press the air release button to release the air in the arm cuff 4 Measurement ends When the measurement is finished the deflation symbol flashes on the display Blood pressure and pulse rate values are displayed Press the air release b...

Page 24: ...to taking the blood pressure measurement 6 Undo the fastener and remove the arm cuff 7 Press the I O button to turn the unit off Note If you forget to turn the unit off it will automatically shut itself off after five minutes Note Self diagnosis of measured results and treatment are dangerous Please follow the instructions of your doctor Diastolic blood pressure Systolic blood pressure Pulse displ...

Page 25: ... research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home This criteria is for home blood pressure measurement For professional office blood pressure measurement criteria please refer to Chapter 9 Some Useful Information about Blood Pressure Systolic Blood Pressure Above 135 mmHg Diastolic Blood Pressure Above 85 mmHg ...

Page 26: ... will be at the same level as your heart 2 Apply the arm cuff to your upper arm The coloured marker should be centred on the inside of your arm and point down the inside of the arm so that the air tube runs down the inside of your forearm and is in line with your middle finger 3 Secure the cuff around your arm using the fabric fastener strip 4 Press the I O button to turn the unit on then pump the...

Page 27: ...d Insert the air tube securely Refer to section 3 2 Arm cuff not applied correctly Apply the arm cuff correctly Refer to section 3 2 Clothing is interfering with the arm cuff Remove any clothing interfering with the arm cuff Refer to section 3 2 Air is leaking from the arm cuff Replace cuff with new one Refer to Chapter 7 The arm cuff was inflated above 299 mmHg Release your hand from the air infl...

Page 28: ...oo low Press the inflation bulb to inflate the arm cuff until the reinflation symbol goes out Or deflate the arm cuff and repeat measurement after checking that the heartbeat symbol has been displayed Refer to section 3 3 Error Display Cause Remedy ...

Page 29: ...ube securely connected into the main unit Make sure that the air tube is connected securely Refer to section 3 2 Air is leaking from the arm cuff Replace the arm cuff with a new one Refer to Chapter 7 Arm cuff deflates too soon The arm cuff is loose Apply the cuff correctly so that it is firmly wrapped around the arm Refer to section 3 2 Cannot measure or readings are too low or too high Has the a...

Page 30: ...re empty Charge the battery or replace the AAA size batteries with new ones Refer to section 2 1 The batteries have been inserted incorrectly Insert the batteries with the correct polarity Refer to section 2 1 Other problems Press the I O button and repeat measurement If the problem continues try replacing the batteries with new ones If this still does not solve the problem contact your OMRON reta...

Page 31: ...pairs of any kind yourself If a defect occurs consult the OMRON distributor or Customer Services as mentioned on the packaging Do not subject the unit to strong shocks or vibrations for example dropping the unit on the floor Do not use volatile liquids to clean the unit THE UNIT SHOULD BE CLEANED WITH A SOFT DRY CLOTH Do not wash the arm cuff or immerse it in water Use a soft moistened cloth and s...

Page 32: ... case Calibration and Service The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature Note Inspec...

Page 33: ...ombination Arm circumference 17 22 cm Small Arm Cuff Arm circumference 17 22 cm Rechargeable battery AC Adapter R CM1 9997578 9 CL1 9996760 3 4997965 1 CSB 4997099 9 Note Should you require a small cuff please ensure that the small cuff and small bulb are used together They may also be purchased as a combination set CS 4997067 0 1140626 7 R Adapter 9997605 0 ...

Page 34: ... Approximately 280 uses when battery is fully charged when inflated to 170 mmHg at 23 C Approximately 300 charge cycles when charged at 23 C 65 RH 2 AAA batteries Approximately 1500 uses with new alkaline batteries when inflated to 170 mmHg at 23 C Operating temperature Humidity 10 C to 40 C 30 to 85 RH Storage temperature Humidity Air Pressure 20 C to 60 C 10 to 95 RH 700 1060 hPa Console Weight ...

Page 35: ...ccording to the European Standard EN1060 Non invasive sphygmomanometers Part 1 General Requirements and Part 3 Supplementary requirements for electromechanical blood pressure measuring systems This OMRON product is produced under the strict quality system of OMRON Healthcare Co Ltd Japan The Core component for OMRON blood pressure monitors which is the Pressure Sensor is produced in Japan for asse...

Page 36: ...erference may result in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to regulate the requirements for EMC Electro Magnetic Compatibility with the aim to prevent unsafe product situations the EN60601 1 2 standard has been implemented This standard defines the levels of immunity to electromag...

Page 37: ... of high blood pressure As a variety of conditions affect blood pressure a single measurement may not be sufficient for an accurate diagnosis Many factors such as physical activity anxiety or the time of day can influence your blood pressure Thus it is best to try and measure your blood pressure at the same time each day to get an accurate indication of any changes in blood pressure Blood pressure...

Page 38: ... on the blood pressure values measured on people in a sitting position in outpatient departments of hospitals Note There is no universally accepted definition of hypotension However those having the systolic pressure below 100 mmHg are assumed as hypotensive 80 180 160 140 130 120 85 90 100 110 Diastolic blood pressure mmHg Systolic blood pressure mmHg Severe hypertension Moderate hypertension Mil...

Page 39: ...ddorp The Netherlands www omron healthcare com Production Facility OMRON DALIAN Co Ltd Economic Technical Development Zone Dalian 116600 China Subsidiary OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 0DG United Kingdom OMRON Medizintechnik Handelsgesellschaft mbH John Deere Str 81a 68163 Mannheim Germany www omron medizintechnik de OMRON SANTÉ FRANCE SAS 14 rue de Lisbonne 93561 ...