30
6. Specifications
Model
HEM-9200T
Display
LCD digital display
Measurement range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Accuracy
Pressure:
±
3 mmHg or 2% of reading
Pulse:
±
5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
IP classification
IP 20
Power source
4 “AA” batteries 1.5V or optional AC adapter (INPUT AC100-240V 50/
60Hz 0.12A)
Battery life
Approximately 1000 measurements (using new alkaline batteries)
Operating temperature /
humidity
50°F to 104°F (10°C to 40°C) / 15 to 90% RH
Storage temperature /
humidity / air pressure
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH /
700 to 1060 hPa
Weight
Monitor : Approximately 10 oz. (290 g)
not including batteries
Arm cuff : Approximately 6 oz. (170 g)
Dimensions
Monitor : Approximately 4 1/4" (w)
×
3 1/8" (h)
×
5 1/2" (l)
(107 mm
×
79 mm
×
141 mm)
Arm cuff : Approximately 5 3/4"
×
23 1/2" (air tube: 29 1/2")
(145 mm
×
594 mm (air tube: 750 mm))
Cuff circumference
9" to 17" (220 to 420 mm)
Contents
Monitor, arm cuff, 4 “AA” batteries, instruction manual,
quick start guide
Applied part
Protection against
electric shock
Internally powered ME equipment (When using only the batteries)
= Type BF
= Class II ME equipment (Optional AC adapter)