background image

IM-HEM-7280T-E-03-02/2017

5344802-4C

Automatic Upper Arm Blood Pressure Monitor

MIT5s Connect (HEM-7280T-E)

Instruction Manual

Summary of Contents for HEM-7280T-E

Page 1: ...IM HEM 7280T E 03 02 2017 5344802 4C Automatic Upper Arm Blood Pressure Monitor MIT5s Connect HEM 7280T E Instruction Manual ...

Page 2: ......

Page 3: ...N Q P R T W V X Y Z U S O A G D C B E F J H I G H K M L AE AD AC H AB H AA ...

Page 4: ...ves a warning signal with the measurement result Important Safety Information 1 1 Know Your Device 4 2 Preparation 7 2 1 Battery Installation 7 2 3 Setting the Date and Time Manually 10 2 2 Using this device with a Smartphone 1 1 Display Symbols 5 1 2 Before Taking a Measurement 6 3 Using the Device 12 3 1 Applying the Arm Cuff 12 3 2 How to Sit Correctly 13 3 3 Taking a Measurement 14 3 4 Using t...

Page 5: ...us drip or blood transfusion Consult your physician before using the device on the arm with an arterio venous A V shunt Do not use the device with other medical electrical ME equipment simultaneously Do not use the device in the area of HF surgical equipment MRI or CT scanner or in an oxygen rich environment The air tube or the AC adapter cable may cause accidental strangulation in infants Contain...

Page 6: ...y result in incorrect measurement results Do not use a mobile phone or other devices that emit electromagnetic fields near the device except when in use for wireless communications This may result in incorrect operation of the device Do not disassemble the monitor or arm cuff This may cause an inaccurate reading Do not use in a location with moisture or a location where water may splash on the dev...

Page 7: ...sed for three months or more General Precautions Do not forcibly crease the arm cuff or the air tube excessively Do not press the air tube while taking a measurement To unplug the air plug pull on the air plug at the connection with the monitor not the tube itself Do not drop the monitor or subject device to strong shocks or vibrations Do not inflate the arm cuff when it is not wrapped around your...

Page 8: ...mbol Average value symbol Systolic blood pressure Diastolic blood pressure Battery symbol low depleted Cuff wrap guide symbol Heartbeat symbol Flashes during measurement Date Time display Morning average symbol Evening average symbol Morning hypertension symbol Irregular heartbeat symbol SYNC symbol Connection symbol OK symbol Deflation symbol Pulse display Memory number Open the rear cover page t...

Page 9: ...plays with your measurement results we recommend you consult your physician Follow the directions of your physician Average Value Symbol The average value symbol is displayed when you press and hold the Memory button for more than 3 seconds The most recent average value appears on the display screen Cuff Wrap Guide Symbol If the cuff was wrapped too loosely it may cause unreliable results If the w...

Page 10: ...nutes before taking a measurement Rest for at least 15 minutes before taking the measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 Measurements should be taken in a quiet place 4 Remove tight fitting clothing from your arm 5 Sit on a chair with your feet flat on the floor Rest your arm on a table so that the arm cuff is at the same level as your heart 6 ...

Page 11: ...r When the depleted battery symbol appears on the display turn the monitor off then replace all batteries at the same time Long life alkaline batteries are recommended The measurement values continue to be stored in memory even after the batteries are replaced The supplied batteries may have a shorter life Dispose of the device components and optional accessories according to applicable local regu...

Page 12: ...nload and install the free OMRON connect app on your smartphone 2 Open the App on your smartphone and follow the set up and pairing instructions 3 Press and hold the connection button for more than 2 seconds The connection symbol and the P flashes on the device s display List of Compatible Smartphones at www omronconnect com devices For details and more information visit www omronconnect com ...

Page 13: ...on the display The monitor will automatically turn off without any operations The date and time setting of your device will automatically sync with your smartphone when paired When Err appears refer to Connection failure in section 4 2 for further details Notes This device can pair to multiple smartphones to support the USER ID function Section 3 3 Any readings previously saved in the blood pressu...

Page 14: ...een paired with a smartphone the year flashes on the Date Time display 2If the device has been paired with a smartphone already the connection symbol and the flash on the device s display Then press the connection button again and the year flashes on the Date Time display Note When the connection symbol and the flashes on the device s display you can now transfer the data refer to page 20 for furt...

Page 15: ...the year once confirmed the month starts flashing Repeat the same steps to change the month day hour and minute 3 Press the START STOP button to turn the monitor off Year Month Day Hour Minute Change Confirm Change Confirm Change Confirm Change Confirm Change Confirm Forward Back ...

Page 16: ... the fabric fastener Notes When you take a measurement on the right arm the air tube will be at the side of your elbow Be careful not to rest your arm on the air tube The blood pressure can differ between the right arm and the left arm and the measured blood pressure values can be different OMRON recommends to always use the same arm for measurement If the values between both arms differ substanti...

Page 17: ...le room temperature Avoid bathing drinking alcohol or caffeine smoking exercising or eating 30 minutes before taking a measurement Sit on a chair with your feet flat on the floor Sit upright with your back straight Sit with your back and arm being supported The arm cuff should be placed on your arm at the same level as your heart c d ...

Page 18: ...ement values in the memory for 2 people using USER ID 1 and USER ID 2 1 Select your USER ID 1 or 2 Push the switch up or down to select USER ID 2 Press the START STOP button The arm cuff starts to inflate automatically START INFLATE DEFLATE COMPLETE Cuff Wrap Guide Symbol If your systolic pressure is more than 210 mmHg After the arm cuff starts to inflate press and hold the START STOP button until...

Page 19: ...Press and hold the START STOP button for more than 3 seconds The USER ID symbol and the Date Time display will disappear 2 Release the START STOP button when the Date Time display turns off The arm cuff will start to inflate automatically DO NOT adjust medication based on measurement results from this blood pressure monitor Take medication as prescribed by your physician Only a physician is qualif...

Page 20: ...er the data before the measurement values are deleted Refer to Using this device with a Smartphone page 8 Flashing 80 sets of measurement results for either user have been stored Lit 100 sets of measurement results for either user have been stored When viewing the measurement value taken without setting the date and time is displayed instead of the date and time To View the Measurement Values Stor...

Page 21: ...etting the date and time the average value is not calculated If there are no measurement values stored in the memory the screen to the right is displayed To View the Weekly Average Value The monitor calculates and displays weekly averages for the measurements taken in the morning and evening within 8 weeks for each user Note The week begins Sunday at 2 00 1 Select your USER ID 1 or 2 ...

Page 22: ...ening average for THIS WEEK appears on the display Press the or button to display the previous weeks Note If there are no measurement values stored in the memory the screen to the right is displayed This week Morning average Last week Morning average 7 weeks ago Morning average Continue to press This week Evening average Last week Evening average 7 weeks ago Evening average 2 weeks ago Morning ave...

Page 23: ...day is calculated for up to three measurements taken within 10 minutes of the last measurement of the evening First measurement of the morning Within 10 minutes 11 59 4 00 Measurements of the morning Up to 3 measurements Morning Blood Pressure Measurement Measurement Measurement Measurement Measurement Measurement Sunday Monday Tuesday Wednesday Thursday Friday Saturday Measurements of the evening...

Page 24: ...e and device differs more than 10 minutes 1 Open the OMRON connect app on your smartphone 2 Pull down to refresh the screen of the OMRON connect app and the data transfer starts If the data transfer is not starting refer to Connection failure Data cannot be sent in section 4 Error Messages and Troubleshooting 3 Confirm with your device if the measurement data is transferred successfully If the dis...

Page 25: ...the button down press the START STOP button for more than 5 seconds 2 Release the button and START STOP button when CLr appears on the display The CLr will appear on the display when your monitor has been initial setting 3 Press the START STOP button to turn the monitor off Note Initial setting of the monitor will not delete the information in the app The monitor will automatically turn off after ...

Page 26: ...leted by USER ID 1 Select your USER ID 1 or 2 2 Press the Memory button while the memory symbol appears 3 While holding the button down press the START STOP button for more than 3 seconds Note You cannot partially delete the values stored in the memory All values for the user you select will be deleted ...

Page 27: ...ohibited On aircraft In hospitals While abroad Make sure that the power has been off before starting the instructions below 1 Press and hold the connection button for more than 10 seconds 2 Release the button when OFF appears on the display 3 Press the START STOP button to turn the monitor off Note The monitor will automatically turn off after 2 minutes ...

Page 28: ...been off before starting the instructions below 1 Press and hold the connection button for more than 10 seconds 2 Release the button when ON appears on the display 3 Press the START STOP button to turn the monitor off Note The monitor will automatically turn off after 2 minutes ...

Page 29: ...on 3 3 If this error continues to appear contact your physician Arm cuff is applied too loosely Apply the arm cuff tighter Refer to section 3 1 The batteries are low You should replace the batteries with new ones ahead of time Refer to section 2 1 The batteries are exhausted You should replace the batteries with new ones at once Refer to section 2 1 Communication failed Refer to Data cannot be sen...

Page 30: ...cuff manually until it is 30 to 40 mmHg above your previous measurement result Refer to section 3 3 The arm cuff was inflated exceeding the maximum allowable pressure and then deflated automatically when inflating the arm cuff manually Do not touch the arm cuff and or bend the air tube while taking a measurement Do not inflate the arm cuff more than necessary Refer to section 3 3 Movement during m...

Page 31: ...o section 3 1 Arm cuff pressure does not rise The air connector is not securely connected into the air jack Make sure that the air tube is connected securely Refer to section 3 1 Air is leaking from the arm cuff Replace the arm cuff with a new one Refer to section 6 1 Arm cuff deflates too soon The arm cuff is loose Apply the arm cuff correctly so that it is firmly wrapped around the arm Refer to ...

Page 32: ...The Bluetooth function on the monitor is not turned on Press the button turn on the Bluetooth function and try sending the data again Pairing registering has not been completed Perform pairing registration Refer to Using this device with a Smartphone in section 2 2 The blood pressure monitor is not ready follow the step in section 2 3 Setting the Date and Time Manually The application on the desti...

Page 33: ...th or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff Changes or modification not approved by the manufacturer will void the user warranty Do not disassemble or attempt to repair the device or components Consult your OMRON retail outlet or distributor Calibration and Service The accuracy of this device has been carefully tested and is designed for a long servic...

Page 34: ... arm cuff Note Do not bend or crease the air tube excessively 3 Place the monitor and the arm cuff in the storage case Do not store the device in the following situations If the device is wet Locations exposed to extreme temperatures humidity direct sunlight dust or corrosive vapours such as bleach Locations exposed to vibrations shocks or where it will be at an angle ...

Page 35: ...jack on the rear side of the monitor 2 Plug the AC adapter into an electrical outlet To disconnect the AC adapter unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor Arm cuff Arm circumference 22 42 cm Small arm cuff Arm circumference 17 22 cm AC adapter HEM RML31 CS2 Small Cuff HEM CS24 AC ADAPTER S 60240HW5SW AC ADAPTER UK 60240H7000SW ...

Page 36: ...00 240V 50 60Hz 0 12A AC ADAPTER UK INPUT AC100 240V 50 60Hz 15VA Battery life Approximately 1000 measurements using new alkaline batteries operating temperature and humidity 23 C 65 RH cuff circumference 25 cm max pressure170 mmHg Applied part Protection against electric shock Internally powered ME equipment When using only the batteries Class II ME equipment Optional AC adapter Operating tempera...

Page 37: ...ices which is the Pressure Sensor is produced in Japan This Product operates in the unlicensed ISM band at 2 4GHz In case this Product is used around the other wireless devices including microwave and wireless LAN which operate same frequency band of this Product there is a possibility that interference occurs between this Product and such other devices If such interference occurs please stop the ...

Page 38: ...d to be observed Do not use mobile cellular telephones and other devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect operation of the device and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in a...

Page 39: ... or other attachments other than the main device itself unless explicitly guaranteed above e Costs arising due to non acceptance of a claim those will be charged for f Damages of any kind including personal caused accidentally or from misuse g Calibration service is not included within the guarantee h Optional parts have a one 1 year warranty from date of purchase Optional parts include but are no...

Page 40: ...ity anxiety or the time of day can influence your blood pressure A single measurement may not be sufficient for an accurate diagnosis Thus it is best to try and measure your blood pressure at the same time each day to get an accurate indication of any changes in blood pressure Blood pressure is typically low in the morning and increases from afternoon to evening It is lower in the summer and highe...

Page 41: ......

Page 42: ...1 2 c d a F L M b ...

Page 43: ...lt Il pacchetto include El paquete incluye Комплект поставки содержит Paketin içindekiler Inhoud van de verpakking ﻋﻠﻰ اﻟﻌﺑوة ﺗﺣﺗوي Automatic Upper Arm Blood Pressure Monitor MIT5s Connect HEM 7280T E Instruction Manual Package includes ...

Page 44: ...pore Industrial Park II Binh Duong Industry Services Urban Complex Hoa Phu Ward Thu Dau Mot City Binh Duong Province Vietnam Subsidiaries Succursales Niederlassungen Consociate Empresas filiales Dochterondernemingen Филиалы Yan Kuruluşlar ...

Reviews: