Page 1: ...3787137 5A Platinum Upper Arm Blood Pressure Monitor Model BP5450 Instruction Manual EN...

Page 2: ...3 Using Your Monitor without a Smart Device 27 3 1 Taking a Measurement 27 4 Using Your Monitor with a Smart Device 31 4 1 Pairing Your Monitor with a Smart Device 31 4 2 Taking a Measurement 32 4 3 D...

Page 3: ...634 4350 before attempting to use this monitor For specific information about your own blood pressure consult with your physician Intended Use This device is a digital monitor intended for use in meas...

Page 4: ...lt in incorrect operation of the monitor and or cause an inaccurate reading DO NOT use this monitor in oxygen rich environments or near flammable gas Consult with your physician before using this moni...

Page 5: ...onsult with your physician before using this monitor if you have had a mastectomy Consult with your physician before using this monitor if you have severe blood flow problems or blood disorders as cuf...

Page 6: ...nment within the temperature specified as operating conditions after it is stored either at the maximum or at the minimum storage temperature For additional information on operating and storage transp...

Page 7: ...nitor Battery Handling and Usage DO NOT insert batteries with their polarities incorrectly aligned ONLY use 4 AA alkaline or manganese batteries with this monitor DO NOT use other types of batteries D...

Page 8: ...ng Cuff AC adapter HEM ADPTW5 4 AA batteries storage case instruction manual quick start guide 1 2 Monitor and Arm Cuff Front of the monitor Display Memory button Morning average button Forward Backwa...

Page 9: ...8 1 Know Your Monitor Left and right side of the monitor Back and bottom of the monitor Air jack AC adapter jack Connection button Date Time Battery compartment g h g h i j i j...

Page 10: ...1 Know Your Monitor 9 EN Arm cuff Upper arm circumference 9 17 22 42 cm Air plug Air tube k k l l...

Page 11: ...od pressure reading Low battery symbol Depleted battery symbol Comparison display Date Time display Pulse display Memory number TruRead symbol Diastolic blood pressure reading High symbol High morning...

Page 12: ...stored in the memory Systolic blood pressure reading Comparison display Displays past readings Refer to section 5 for more information Pulse display Memory number Pulse rate appears after the measure...

Page 13: ...t reading Up to 100 readings can be stored in the internal memory of your monitor per user Low battery symbol Flashes when batteries are low Depleted battery symbol Appears when batteries are depleted...

Page 14: ...ol Appears when the morning weekly average is 130 80 mmHg or above OK symbol Flashes when your monitor is connected to your smart device or readings are transferred successfully Movement error symbol...

Page 15: ...25 less or 25 more than the average rhythm detected while your monitor is measuring blood pressure Irregular heartbeat symbol may be displayed when the monitor detects arm movements Refer to section...

Page 16: ...ur physician BP Category Systolic BP Diastolic BP Normal 120 mmHg and 80 mmHg Elevated 120 129 mmHg and 80 mmHg Hypertension Stage 1 130 139 mmHg or 80 89 mmHg Stage 2 140 mmHg or 90 mmHg Individuals...

Page 17: ...time manually or refer to sub section 4 1 to pair your monitor to set the date and time automatically Note When the symbol flashes on the display it is recommended you replace batteries To replace ba...

Page 18: ...te and time each time you unplug and plug back the AC adapter The readings will not be deleted 1 Insert the AC adapter plug into the AC adapter jack on the right side of your monitor 2 Plug the AC ada...

Page 19: ...t US CAN app Refer to sub section 4 1 As soon as batteries are correctly installed the year flashes on the display of your monitor 1 If nothing is displayed press the button rear surface of the monito...

Page 20: ...o store the setting TruRead mode setting is displayed 3 seconds after the minutes have been set in this sequence Refer to sub section 2 4 Note If you need to reset the date and time replace batteries...

Page 21: ...e is set to oFF by default 1 Set or bypass the date and time settings Follow the steps in sub section 2 3 and set or bypass the date and time settings After it is completed the symbol appears on the d...

Page 22: ...ton to change the interval It can be set 15 30 60 or 120 seconds 5 Press the button to confirm Your monitor will automatically turn off in 3 seconds To take a measurement in TruRead mode refer to Usin...

Page 23: ...Taking measurements in the morning and in the evening is recommended Remember to have a record of your blood pressure and pulse readings for your physician A single measurement does not provide an acc...

Page 24: ...od pressure can differ between the right arm and the left arm and the measured blood pressure readings can be different OMRON recommends to always use the same arm for measurement If the readings betw...

Page 25: ...dge of the arm cuff should be 0 5 inch 1 to 2 cm above the inside elbow The air tube is on the inside of your arm and aligned with your middle finger 3 Make sure that the air tube is positioned on the...

Page 26: ...N To take a measurement on your right arm When you take a measurement on your right arm the air tube should run along the side of your elbow along the bottom of your arm Be careful not to rest your ar...

Page 27: ...table chair with your back and arm supported Keep your feet flat and your legs uncrossed The arm cuff should be placed on your arm at the same level as your heart with the arm resting comfortably on a...

Page 28: ...the monitor and or cause an inaccurate reading Remain still and DO NOT talk while taking a measurement 1 Press the START STOP button All symbols appear on the display before starting the measurement...

Page 29: ...te You may need to increase the wait time depending on your individual physiological characteristics Using the TruRead Mode Set the TruRead mode on to use the TruRead mode Refer to sub section 2 4 1 S...

Page 30: ...art The interval you selected during TruRead settings will appear on the display After the third measurement is complete the average for the 3 measurements appears on the display To view readings in T...

Page 31: ...the date time display will disappear 2 Release the START STOP button when the user ID and date time display turn off The arm cuff automatically starts to inflate If your systolic pressure is more than...

Page 32: ...ll the free OMRON connect US CAN app onto your smart device If you already have the OMRON connect US CAN app and have created your account open the app and add your new monitor 3 Open the app and foll...

Page 33: ...tion in the app OMRON connect US CAN is the only app that we recommend to use with your monitor to transfer data correctly 4 2 Taking a Measurement Refer to sub section 3 1 to start a measurement Open...

Page 34: ...s For further information on potential restrictions refer to documentation on the Bluetooth usage by the FCC ISED Bluetooth is enabled as default 1 When your monitor is off press and hold the button r...

Page 35: ...mparison display on the left Note If the reading is high refer to sub section 1 3 the symbol appears If there are no readings stored in the memory the screen to the right is displayed If the memory is...

Page 36: ...gs with the TruRead symbols are the averages for the consecutive 3 measurements To view the individual readings press the button rear surface of the monitor while the TruRead average is displayed 1 Wh...

Page 37: ...our monitor to the correct date and time before taking a measurement to view the average readings 1 Select your user ID 2 When your monitor is off press and hold the button for more than 3 seconds The...

Page 38: ...age readings Morning Weekly Average Calculation This is the average for the measurements taken during the morning 4 00 am 9 59 am between Sunday and Saturday The 2 or 3 readings taken within the first...

Page 39: ...the left If the morning weekly average is high refer to sub section 1 3 the or symbol appears 3 Press the or button repeatedly to scroll through the previous readings stored in the memory Your monito...

Page 40: ...s the latest reading will appear with a past reading 3 While holding the button down press and hold the START STOP button for more than 3 seconds The display to the right flashes 4 While holding the b...

Page 41: ...ettings To delete all the information stored in your monitor follow the instructions below Make sure that your monitor is turned off 1 While holding the button rear surface of the monitor down press a...

Page 42: ...r the completion the year flashes on the display Refer to sub section 2 3 to set the date and time manually or refer to sub section 4 1 to pair your monitor to set the date and time automatically Reve...

Page 43: ...r measurement Refer to sub section 2 6 Air is leaking from the arm cuff Replace the arm cuff with a new one Refer to section 9 appears or a measurement cannot be complete after the arm cuff inflates Y...

Page 44: ...ent Remain still and do not talk during a measurement appears The pulse rate is not detected correctly Apply the arm cuff correctly then take another measurement Refer to sub section 2 6 Remain still...

Page 45: ...ART STOP button to cancel pairing and turn your monitor off flashes The monitor is ready to transfer your readings to the smart device Open the OMRON connect US CAN app to transfer your readings flash...

Page 46: ...tress time of day and or how you apply the arm cuff may affect your blood pressure Review sub sections 2 5 2 7 and section 3 Any other communication issue occurs Follow the instructions shown in the s...

Page 47: ...nd other components in the storage case when not in use Store your monitor and other components in a clean safe location 1 Remove the arm cuff from the monitor Caution To unplug the air plug pull on t...

Page 48: ...or immerse your monitor and arm cuff or other components in water Do not use gasoline thinners or similar solvents to clean your monitor and arm cuff or other components 8 4 Disposal Dispose of your...

Page 49: ...ormation visit our website at OmronHealthcare com Small D Ring Cuff Wide Range D Ring Cuff Arm circumference 7 9 17 22 cm Arm circumference 9 17 22 42 cm Model HEM CS24 B HEM CS24 BZ Model HEM RML31 B...

Page 50: ...nergy Wireless communication Frequency range 2 4 GHz 2400 2483 5 MHz Modulation GFSK Effective radiated power 20 dBm Operation mode Continuous operation IP classification Monitor IP20 AC adapter IP21...

Page 51: ...ed against vertically falling water drops which may cause issues during a normal operation Weight Monitor Approximately 16 6 oz 470 g not including batteries Arm cuff Approximately 6 oz 170 g Dimensio...

Page 52: ...this product is used near other wireless devices such as microwave and wireless LAN which operate on the same frequency band as this product there is a possibility that interference may occur If inter...

Page 53: ...ecifications Pictogram on the Product or Types of marks used to indicate the correct positioning of the artery for the upper arm Identifies the type of cuff that is compatible with the blood pressure...

Page 54: ...erference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interfe...

Page 55: ...with FCC ISED radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency RF Exposure Guidelines and RSS 102 of the ISED radio frequency RF Exposure rules as...

Page 56: ...nd the return shipping and handling fee that may apply Enclose the original printed receipt Include a letter with your name address phone number and description of the specific problem Pack the produc...

Page 57: ...may have other rights that vary by jurisdiction Because of special local requirements some of the above limitations and exclusions may not apply to you FOR CUSTOMER SERVICE Visit our web site at For...

Page 58: ...ocumentation in accordance with this EMC standard is available at OmronHealthcare com emc Refer to the EMC information for BP5450 on the website The Bluetooth word mark and logos are registered tradem...

Page 59: ...58 Notes...

Page 60: ...ARE Co Ltd 53 Kunotsubo Terado cho Muko Kyoto 617 0002 JAPAN Distributed by OMRON HEALTHCARE INC 1925 West Field Court Lake Forest IL 60045 U S A For USA OmronHealthcare com For Canada OmronHealthcare...

Reviews: