© 2011 Omnicell, Inc.
1-, 2-, 3-Cell Color Touch Installation and Service Guide/
67-2014 Rev G
Switchpanels
4-3
Removal Procedure
From a Supply Drawer
1. Disconnect the drawer ribbon cable from the SPC board in the wireway.
Figure 4-3.
Disconnect the ribbon cable from the SPC Board in the Wireway
2. Remove the two
on the bottom of the supply drawer using a T15 Torx screwdriver.
Figure 4-4.
Remove the screws securing the drawer to the mounting bracket
3. Remove the supply drawer and its casing from the dispenser unit.
Caution:
If pliers are used to remove the drawer ribbon cable, ensure that the connection is removed slowly
and that the pliers are pulling both the cable and the connection. If the connector on the end of the ribbon
cable comes free from the ribbon cable, it might cause a short, damaging the wireway or the PowerCom
board.
Note:
If the supply drawer is to be located directly above a pharmacy drawer, it is recommended that the
screws be re-installed from above of the bracket.