OMNI 4000/7000 Operations and Maintenance Guide
– Rev F
6-6
Front Panel Operations
6.3.2 Edit Prove Transducers
To edit proving process input variables, follow these instructions:
1.
a. In the
Home
menu, use the arrow
navigation keys to go to and
select
Prove
.
b. Press the
OK
key.
2.
a. Use the
Right
or
Left
arrow keys to
select the Meter Run (at the top of the
screen) for which you need to edit the
proving process input variables.
b. Press the
EditXdcr
function key.
3.
This is the
Prove Transducers
screen,
which lists the editable process input
variables for the prove.
a. Use the
Up
or
Down
arrow keys to
choose the process input variable you
wish to edit.
b. Press
OK
.
4.
For the process input variables
temperature, pressure and density, you
can edit the
Alarm Limits
and
Override
values.
a. Use the arrow keys to select the
variable you wish to edit
b. Press
OK
to access the field and
make changes.
c. Press
OK
or the
Accept
function key
to save your changes.
You can also access this screen
directly by selecting
Operate
on
the
Home
screen, selecting
Edit
Transducers,
and using the arrow
keys to navigate past the Meter
Runs and Stations to the Process
Variables (
PV
) screen.