CPC 100 V1.41
2 - 16
Context-dependent menu keys
Soft-touch keys that directly invoke specific commands associated with the
currently selected control of the test card and view.
Since their actual functions depend on the selected view, test mode, test card
and selected UI element, the menu keys do not hold any specific text. The
individual key function is displayed on the monitor with a labelled button symbol
right next to the menu key.
I/O (test start/stop)
If the test parameters are set in the software’s Test Card View, the Emergency
Stop button is released and the safety key lock is set to the "enable" position,
pressing this push-button enables
CPC 100
outputs and starts a test.
Pressing this push-button a second time stops the test.
Tab selector
If a test procedure comprises more than one test card, use these keys to change
between the single test cards.
Note:
Stopping a test
does not
shut off
the
CPC 100
outputs
instantaneously
. First, the currently running test sequence finishes,
then the test execution is stopped.
Most test cards finish the running test sequence with a predefined
ramp function.
Therefore
,
in a hazardous situation never press "Stop Test". Instead,
use
Emergency Stop
.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...