CPC 100 V1.41
2 - 6
Functional Components in Detail
AC OUTPUT
Programmable AC current and voltage outputs.
Figure 5:
Functional group
AC OUTPUT
1. The front panel output
6A/130V
operates as either a 0...6A AC or a
0...130V AC output. The output range and the actual values are set by the
software.
In current mode 0...6A, the maximum output voltage is 65V (390VA).
In voltage mode 0...130V, the maximum output current is 3A (390VA).
The
6.3A T
(6.3 Amps slow-acting wire fuse 5x20mm) protects both the
AC OUTPUT
in 6A current mode and the
DC OUTPUT
.
The
3.15A T
(3.15 Amps slow-acting wire fuse 5x20mm) protects the
AC OUTPUT
in 130V voltage mode and in the 3A AC current mode.
2. The
800A
output sockets are situated at the left-hand side of
CPC 100
as
shown on page 2-3.
They provide an AC current of 0...800A at a voltage of 6.1...6.5V. The actual
output power (4880VA max.) is range-dependent. For details refer to ”High-
Current and High-Voltage Outputs” on page 16-5.
The current that flows from this output is measured internally by
CPC 100
,
and displayed on the respective test card.
3. The
2kV
output sockets are situated at the left-hand side of
CPC 100
as
shown on page 2-3.
Lit LED indicates that
output is selected.
Summary of Contents for CPC 100
Page 12: ...CPC 100 V1 41 x...
Page 28: ...CPC 100 V1 41 1 16...
Page 90: ...CPC 100 V1 41 3 14...
Page 194: ...CPC 100 V1 41 6 30...
Page 250: ...CPC 100 V1 41 8 32 Figure 24 Settings of Amplifier test card for this example use case...
Page 258: ...CPC 100 V1 41 9 4...
Page 264: ...CPC 100 V1 41 10 6...
Page 282: ...CPC 100 V1 41 12 10 Figure 9 Saving tests with the CPC Editor...
Page 284: ...CPC 100 V1 41 12 12...
Page 312: ...CPC 100 V1 41 14 12...
Page 316: ...CPC 100 V1 41 15 4...
Page 350: ...CPC 100 V1 41 16 34...
Page 372: ...OMICRON Contact Addresses 22...