CP Strobe Warning Lamp
45
OMICRON
Figure 19:
CP Strobe
Warning Lamp connected to the main device (here
CIBANO 500
) with the
CP Strobe
connection cable
7.4
Testing
To test the
CP Strobe
Warning Lamp:
1. Disconnect all test leads and cables – except the
CP Strobe
connection cable – from the main
device. Do not connect any cables during the test of the
CP Strobe
Warning Lamp.
2. Release the
Emergency Stop
button on the front panel of the main device.
3. Create an arbitrary manual test as described in the User Manual of the main device.
NOTE!
See one or more of the following manuals:
•
CIBANO 500
User Manual
•
CPC PTM
User Manual
•
TESTRANO 600
User Manual
4. Execute the created test by clicking
Start
in the
Primary Test Manager
workspace and pressing the
Measurement Start/Stop
button on the front panel of the main device.
5. Now, the
CP Strobe
Warning Lamp should flash while the test is running.
6. Press the main device’s
Measurement Start/Stop
button.
7. Press the main device’s
Emergency Stop
button.
Safety Connector Dongle
Main device
CP Strobe
Warning Light
CP Strobe
connection cable