OMICRON
29
Technical data
The accompanying diagram shows the load limit curve for DC voltages. For AC voltages, a maximum
power of 2000 VA is achieved.
Load limit breaking capacity curve for binary output relays with DC voltages
4.4.6
DC supply (AUX DC)
DC supply (AUX DC)
Voltage ranges
0 ... 66 V
DC
(max. 0.8 A)
0 ... 132 V
DC
(max. 0.4 A)
0 ... 264 V
DC
(max. 0.2 A)
Power
Max. 50 W
Accuracy
1
1. Percentage is with respect to each range’s full-scale.
Typical
Guaranteed
Error <2 %
Error <5 %
Resolution
<70 mV
Connection
4 mm sockets on front panel.
Short-circuit protection
Yes
Overload indication
Yes
Insulation
Reinforced insulation from power supply and all SELV interfaces.
DC current (A DC)
300
200
100
50
40
30
20
10
0.1 0.2
0.5
1
2
5
10
20
Resistive load
L/R = 50 ms
DC voltage (V DC)