Error and Warning Messages
12
Error and Warning Messages
001.xxx
Error [001] No valid CT software!
Reason:
No valid software in the device Flash memory or incorrect checksum.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
Omicron
directory and switch the
CT Analyzer
off and on again.
002.xxx
Error [002] Can't open file!
Reason:
The firmware cannot read the file
CTAnalyzer.bin
from the Compact Flash
card since the CF card or the file is missing.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
Omicron
directory and switch the
CT Analyzer
off and on again.
003.xxx
Error [003] Download error!
Reason:
The downloaded software is corrupt.
Solution:
Insert a Compact Flash card with valid software (
CTAnalyzer.bin
) in the
Omicron
directory and switch the
CT Analyzer
off and on again.
100.xxx
Warning [100.xxx] CT resistance > 3000 Ohms.
Reason:
The resistance measured during primary or secondary winding resistance
measurement is > 3000
Ω
.
Solution:
The test cannot be continued. Check the connections to the CT. If the
connections are ok, the CT resistance is > 3000
Ω
. Such CTs cannot be tested
using the
CT Analyzer
.
101.xxx
Warning [101.xxx] Timeout during measurement. No constant winding resistance can
be determined within 10 sec.
Reason:
The
CT Analyzer
cannot determine a constant winding resistance within
10sec. during primary or secondary winding resistance measurement.
Solution:
The test cannot be continued. Check the connections to the CT.
110.xxx
Warning [110.xxx] The impedance of the CT is too high.
Reason:
The impedance of the CT or the load is too high to reach the minimum toggle
frequency of 0.2Hz / 0.8Hz.
Solution:
The test cannot be continued. Check the connections to the CT. If the
connections are ok, it is not possible to test this CT due to its too high
inductance.
111.xxx
Warning [111.xxx] Inductance of load too low.
Reason:
Measurement frequency is too high and cannot be decreased because of too
low CT inductance.
Solution:
The test cannot be continued. Check the connections to the CT. If the
connections are ok, it is not possible to test this CT due to its too low
inductance.
112.xxx
Warning [112.xxx] Measurement error, reduce noise level.
Reason:
It is not possible to obtain stable measurement results due to incorrect wiring,
wiring short-circuit, connected load or external interferences.
Solution:
Check the wiring. The primary side of the CT must not be short-circuited.
Disconnect the primary side of the CT from the transmission lines, connect one
primary side terminal to PE and open the other one. Use the delivered original
coax cables for measurement (see section 2.7.2 ”Noise Reduction
Techniques” on page 43). Make sure that no burden is connected to the
secondary side of the CT. This error may also indicate a winding short circuit
of the CT.
Summary of Contents for CT Analyzer
Page 1: ...CT Analyzer CT SB2 User Manual...
Page 3: ...CT Analyzer User Manual...
Page 4: ...CT Analyzer User Manual 2 Manual Version CTAnalyzerUM AE 3...
Page 39: ...37 Setup and Connection Figure 2 15 Measurement setup for delta compensation Ratio 1 L2 L3 L1...
Page 50: ...CT Analyzer User Manual 48...
Page 60: ...CT Analyzer User Manual 58...
Page 80: ...CT Analyzer User Manual 78...
Page 162: ...CT Analyzer User Manual 160...
Page 172: ...CT Analyzer User Manual 170...
Page 188: ...CT Analyzer User Manual 186...
Page 194: ...CT Analyzer User Manual 192...
Page 195: ...CT SB2 User Manual...
Page 196: ...CT SB2 User Manual 2 Manual Version CTSB2 AE 3...
Page 204: ...CT SB2 User Manual 10...
Page 206: ...CT SB2 User Manual 12...
Page 211: ...17 Hardware 1 x Cable bag Bag for cables and accessories Accessories Description...
Page 212: ...CT SB2 User Manual 18...
Page 226: ...CT SB2 User Manual 32...
Page 242: ...CT SB2 User Manual 48...
Page 248: ...CT SB2 User Manual 54...
Page 250: ...CT SB2 User Manual 56...
Page 252: ...CT SB2 User Manual 58...