8.4.5 Automatic deletion of calibration
In order to avoid false measurements or an invalid calibration, the
Bode Analyzer Suite
automatically
deletes calibration as soon as a setting makes the calibration invalid. The following lists show which
settings will delete a calibration when they are changed.
Deleting User-Range calibration
User-Range calibration is automatically deleted when one of the following settings is changed:
•
Measurement frequencies (Start Frequency, Stop Frequency, Center Frequency, Span)
•
Sweep Mode (Linear or Logarithmic)
•
Number of Points (in the frequency sweep)
•
Channel termination (1 MΩ or 50 Ω)
•
Reference / receiver connection (Internal or External)
•
External probe factor
Deleting Full-Range calibration
Full-Range calibration is automatically deleted when one of the following settings is changed:
•
Channel termination (1 MΩ or 50 Ω)
•
Reference / receiver connection (Internal or External)
•
External probe factor
8.4.6 Saving calibration data
User-Range and Full-Range calibration data is saved to the .bode3 file.
The calibration data (correction values) as well as the calibration states (activated or not) are stored to
the file. When opening a bode-file the
Bode Analyzer Suite
attempts to load the external calibration
data and calibration states from the file.
The
Bode Analyzer Suite
attempts to load the calibration from the bode-file even when the
calibration has been performed with a different
Bode 100
device.
Hardware incompatibilities between Bode 100 R1 and Bode 100 R2
•
Full-Range calibration from
Bode 100 R1
is not compatible to Full-Range calibration of
Bode 100 R2
. Full-Range calibration will be deleted automatically when opening a bode-file
created with a different hardware revision.
•
User-Range calibration is not compatible in the frequency range from 3 kHz to 30 kHz. A User-
Range calibration that contains frequencies in that range will be deleted automatically when
opening a bode-file created with a different hardware revision.
Calibration
OMICRON Lab
91
Summary of Contents for Bode 100
Page 1: ...Smart Measurement Solutions Bode 100 User Manual...
Page 24: ...Figure 6 3 Collapsing the measurement and trace configuration areas 24 OMICRON Lab...
Page 45: ...Figure 7 3 Context menu of chart Measurement types and applications OMICRON Lab 45...
Page 83: ...Open B WIC Open B SMC Short B WIC Short B SMC Load B WIC Load B SMC Calibration OMICRON Lab 83...