Calibrating Open/Short/Load:
Open calibration.
Short calibration.
Note that the inductance of the short connection
is assumed to be zero. Short Delay Time is 0 s by
default.
Load calibration.
Default Load Resistor value setting is 50 Ω.
For more details and practical examples on how to perform calibration in the Voltage/Current
measurement mode, please refer to the corresponding application notes on
8.4 Further calibration information
8.4.1 Difference between Full-Range and User-Range calibration
Full-Range
calibration and
User-Range
calibration differ only in the
frequencies
that are used to
measure the correction factors.
Full-Range
calibration measures the correction factors over the
"full" frequency range
of the
instrument at factory-predefined frequencies.
User-Range
calibration measures the correction factors at
exactly
the same
frequency
range and
frequency points that are used in the
measurement
currently configured by the user.