1.7 Quick check of attenuators
By having a look at the level indicators in the lower right corner of the status bar a quick check of the
input attenuators can be performed. This is just a qualitative check that proves that the attenuators do
their job in general.
Measurement setup:
Connect the Source of Bode 100 to the two inputs using the cables and the T-BNC Adapter delivered
with Bode 100.
Figure 32: quick check attn. - measurement setup
Equipment setup:
Please select the
measurement type “Gain / Phase” and apply the following settings in the hardware
setup:
Figure 33: quick check attn. - measurement type
Figure 34: quick check attn.
– open and change hardware setup