Omicron Lab B-AMP 12 User Manual Download Page 1

B-AMP 12

Power amplifier for Bode 100 

User Manual 

Distributed by:

dataTec

  ▪  Ferdinand-Lassalle-Str. 52  ▪  72770 Reutlingen  ▪  Tel. 07121 / 51 50 50  ▪  Fax 07121 / 51 50 10  ▪  [email protected]  ▪  www.datatec.de

Summary of Contents for B-AMP 12

Page 1: ...B AMP 12 Power amplifier for Bode 100 User Manual Distributed by dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 10 info datatec de www datatec de...

Page 2: ...tion and or storage in electronic data processing systems requires the explicit consent of OMICRON electronics Reprinting wholly or in part is not permitted The product information specifications and...

Page 3: ...for household use After use the test set cannot be disposed of as household waste For customers in EU countries incl European Economic Area OMICRON test sets are subject to the EU Waste Electrical an...

Page 4: ...aller or equal to 1 Vrms in the frequency range from DC 50 MHz B AMP 12 has been especially designed for the use in laboratory and manufacturing environments Safety symbols used In this manual the fol...

Page 5: ...is a SELV device Safety Extra Low Voltage in accordance to IEC 60950 also known as protection class III equipment Do not apply hazardous voltage levels 50 VDC or 25 VAC to any input or output of B AMP...

Page 6: ...1 output connector BNC plug to measure the internal voltage attenuated Power supply connector Supply the B AMP 12 with 10 V DC to 36 V DC Channel 2 output connector BNC plug to measure the output vol...

Page 7: ...cal setup where B AMP 12 is used to amplify the output signal of the Bode 100 vector network analyzer to measure S21 of a 2 port DUT using up to 25 dBm signal level Figure 4 2 shows a typical setup us...

Page 8: ...tput amplitude 25 dBm 11 3 VPP into 50 Output impedance 50 Output return loss 30 dB Output DC offset 10 mV input and output terminated with 50 Damage reverse power 1 W Input to CH1 attenuation 14 5 dB...

Page 9: ...rature 5 C 40 C 41 F 104 F Storage temperature 35 C 60 C 31 F 140 F Relative humidity 20 80 non condensing Table 5 4 Mechanical data Characteristic Rating Dimensions with connectors 138 mm x 75 mm x 3...

Page 10: ...b com At our technical support hotline you can reach well educated technicians for all your questions Competent and free of charge Make use of our technical support hotlines Americas 1 713 830 4660 or...

Page 11: ...www omicron lab com OMICRON Lab 2017 B AMP 12 User Manual English All rights reserved ENU11880501 Distributed by dataTec Ferdinand Lassalle Str 52 72770 Reutlingen Tel 07121 51 50 50 Fax 07121 51 50 1...