Maintenance and cleaning
Series TCL-3M165E2
- 52 -
Measuring insert (Adapter sleeves and calibration inserts):
During operation, small quantities of metal dust get created. It can result in the measuring
insert getting stuck in the calibration block.
Pull the measuring insert out of the calibration block with the help of the sleeve exchange
tool.
Clean the measuring insert and the calibration block at regular intervals.
Before a prolonged shutdown of the calibrator, remove the measuring insert from the
calibration block.
Tank (Micro bath calibrator):
Before cleaning, the tank must be drained as much as possible with the drain syringe.
When doing so, follow the corresponding instructions in the safety data sheet of the
calibration liquid used.
Distilled Water:
Remove the sensor cage from the tank.
Remove the magnetic stirrer with the help of the magnetic lifter.
Dry the tank, the sensor cage, the magnetic stirrer and the drain syringe thoroughly.
Silicone oil:
Remove the sensor cage from the tank.
Remove the magnetic stirrer with the help of the magnetic lifter
Clean the cage, the magnetic stirrer and the tank with water to which a generous amount
of rinsing liquid has been added.
Remove the cleaning water to the maximum extent possible using the drain syringe.
Dry the tank, the sensor cage, the magnetic stirrer and the drain syringe thoroughly.
8.3 Recalibration
The calibrator is adjusted and tested with measuring equipment in accordance with
recognized national standards prior to delivery.
The calibrator should, depending on the application situation, be inspected at appropriate
intervals on the basis of ISO 10012. We recommend you to return the calibrator to OMEGA
ENGINEERING at intervals of max. 12 months or approx. 500 operating hours for
recalibration and readjustment.