HHF1000 SERIES

Handheld Air Velocity/ 

Temperature Meter with Wireless 

PC Communication Option

e-mail: [email protected] 

For latest product manuals: 

www.omegamanual.info

Shop online at 

omega.com

 SM

User’s Guide

Summary of Contents for HHF1001A

Page 1: ...HHF1000 SERIES Handheld Air Velocity Temperature Meter with Wireless PC Communication Option e mail info omega com For latest product manuals www omegamanual info Shop online at omega com SM User sGuide ...

Page 2: ...specifications without notice omega com info omega com Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide ...

Page 3: ...ion 1 1 Section 2 Installation 2 1 Section 3 Main Operation 3 1 Section 4 PC Interface 4 1 Section 5 Wireless PC Communications 5 1 Section 6 Specifications 6 1 Section 7 Approvals Regulatory Compliance 7 1 HHF1000 SERIES Handheld Air Velocity Temperature Meter ...

Page 4: ...4 2 Typical Data File 4 2 4 3 Settings Menu 4 3 5 1 UWTC REC1 Wireless Receiver 5 1 5 2 General and Mounting Dimensions 5 2 5 3 FCC Warning Label 5 2 5 4 Establishing Communications With HHF1000 Screen 5 3 5 5 Options Menu Screen 5 4 5 6 TC Central Main Menu Screen 5 5 5 7 Y Axis Configuration Screen 5 6 5 8 Data Plotting Logging Menu Selection Screen 5 6 5 9 Air Velocity Temperature Plot In Real ...

Page 5: ...nd kilometer per hour Km h The air temperature is displayed in ºF ºC The volume air flow is displayed in Cubic Feet per minute CFM and Cubic meter per minute CMM The sensor probe is 12 long as standard The 304 Stainless steel sensor tubing is provided with inch marks for ease of insertion depths The sensor probe comes in two different versions as follows Fixed 12 inches long Probe Telescopic probe...

Page 6: ...asurement In addition air carrying dust or oil such as found in blower compressor systems that utilize oil can lead to coating of the sensor and thus inaccurate readings Refer to the Maintenance section page 4 3 for information on cleaning the sensor The HHF1000 is a bi directional device meaning the air flow in the forward or reverse direction provides the same readings Introduction 1 1 2 ...

Page 7: ...5 diameters of straight pipe regardless of the downstream obstruction 4 Align the sensor probe with the air flow Make sure the air flow is perpendicular to the sensor window The score line on the sensor tubing is another way of aligning the sensor to the flow stream The score line starts from the center of the sensor window and as a result it can be aligned properly 5 One way of installing the sen...

Page 8: ... A A A A A A A 20D 15D Elbow Tee Branch Pipe 25D 15D Elbow 2 planes 25D 15D Long radius bends 30D 25D 15D 15D Elbow Long radius bends 40D 35D 15D 15D Elbow Long radius bends 20D 15D Contracting Pipe 40D 20D Expanding Pipe Recommend Meter Be Installed Upstream Regulating reducing valves Ball check valves Shut off valves V aried Section Va lves Fittings in Tw o Planes All Fittings in Same Plan e Not...

Page 9: ...CONNECTION PROBE INPUT USB PORT DC POWER INPUT 9V 100mA FCC ID OUR XBEEPRO IC 4214A XBEEPRO This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not cause harmful interference 2 This device must accept any interference received including interference that may cause undesired operation FC C Made in U S A RUBBER BOOT RUBBER BOOT 12...

Page 10: ... 914 4 36 inches RX TX MODE SET F C HHF1000 SERIES OMEGA AIR VELOCITY TEMPERATURE METER USB PORT DC POWER INPUT 9V 100mA USB PORT DC POWER INPUT 9V 100mA AIR VELOCITY TEMPERATURE SENSOR FIXED PROBE MOUNT HHF1000 SERIES 1 4 OD 304 STAINLESS STEEL TUBING WITH INCH MARKING AIR FLOW TO BE PERPENDICULAR TO THE SENSOR WINDOW AIR FLOW SCORE LINE FOR FLOW ALIGNMENT FPM F FPM F TELESCOPIC AIR TEMP PROBE 30...

Page 11: ...WTC REC1 and be able to display and graph data on the TC central software application Changing Display Engineering Units You can change the air velocity Engineering unit display from Feet per minute FPM to meter per second m s miles per hour MPH and kilometers per hour Km h from the keypad Press SET key You can change the Temperature display from ºF to ºC or vise versa by pressing the Down key Tur...

Page 12: ...AL Alarm Set Point Alarm Set Point Set Point Mode Icon When Enabled Display LCD Go to Cross ____________ Increment Decrement Response time Sec Area Mode Response Time Response Time Display Cross Go to Atmos ____________ Increment Decrement Sectional Area Sq In Press Mode Cross Sec Area Cross Sec Area Display AtPr Go to Power Increment Decrement mmHg Icon Shutoff Mode ____________ Atmospheric Atmos...

Page 13: ... DISPLAY MODE PRESS MODE PRESS MODE PRESS MODE PRESS MODE PRESS MODE PRESS MODE FPM C PRESS MODE F m s F MPH F Km h PRESS F C PRESS SET REAL TIME MODE m s REAL TIME MODE MPH PRESS SET PRESS SET REAL TIME MODE Km h PRESS SET REAL TIME MODE C PRESS TO CHANGE CROSS SECTIONAL AREA PRESS TO CHANGE RESPONSE TIME PRESS TO CHANGE ATMOSPHERIC PRESSURE PRESS TO CHANGE LOW ALARM SET POINT PRESS TO CHANGE HIG...

Page 14: ...tion with the HHF1000 meter and a Go button will show up on the Main window Clicking the Go button will start the data logging session If the application can not establish communication it will show an error message box and the Find button will flash Please check for the following The meter is connected to a USB port of the PC Go to Settings menu and check the COM port number Make sure you are usi...

Page 15: ...u can save the logged data into a data file This file can be imported into Excel spread sheet for further analysis Figure 4 2 shows a typical data file You can save up to 144 000 sets of velocity and temperature data Figure 4 2 Typical Data File Screen ...

Page 16: ...tory viewer and Maximize start up When the Save to File is selected Default the program asks if you would like to save the data to a file when you stop the data logging Figure 4 2 shows a typical data file saved Figure 4 3 Settings Menu Screen Maintenance Except for intermittent removal of the sensor from the line for cleaning there is no routine maintenance for the HHF1000 If the sensor probe bec...

Page 17: ...ll only blink when the receiver is connected to your PC and you initialize your measurement software After the receiver establishes communication with the program the light will no longer blink Note this may happen very fast and will not be noticeable Receiver RX red indicator light The red indicator light marked RX on the front of the receiver will blink each time the receiver receives incoming d...

Page 18: ...dress and data transmission interval using this feature Wireless PC Communications 5 5 2 ANTENNA USB I O 9 24 Vdc OMEGA ENGINEERING INC Stamford CT 06907 UWTC SERIES WIRELESS RF RECEIVER 2 4 GHz omega com Made in U S A TX RX SB PWR This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not cause harmful interference 2 This device m...

Page 19: ...ovided Run the End device configuration software The software will recognize the transmitter and allows the settings to be changed You can set the data transmission interval from 1 to 60 seconds You can have up to 48 wireless transmitters sending data to one wireless receiver Figure 5 4 shows the End Device Configuration software establishing communication with HHF1000 Figure 5 5 shows the Options...

Page 20: ...or wireless transmission 1 If the sensor probe is disconnected or any sensor failure the red LED blinks 3 times every 15 seconds 2 If there is an RF communication failure the red LED blinks once every 15 seconds 3 If there is a sensor failure and RF communication failure at the same time the red LED blinks 4 times every 15 seconds Wireless PC Communications 5 5 4 NOTE NOTE ...

Page 21: ...xis for air velocity and temperature by clicking the options button Figure 5 7 shows the configuration of the Y Axis You can chart and log data in real time Simply click on the chart tab from the main menu After configuring the Y Axis click on the start button from the chart tab menu It will take you to the Data Plotting Logging Menu selection as shown in Figure 5 8 In this menu you can select cha...

Page 22: ...Figure 5 7 Y Axis Configuration Screen Figure 5 8 Data Plotting Logging Menu Selection Screen Wireless PC Communications 5 5 6 ...

Page 23: ...Figure 5 9 Air Velocity Temperature Plot In Real Time Screen Figure 5 10 Data Log Screen Wireless PC Communications 5 5 7 ...

Page 24: ...d Temperature Values Maximum Minimum 250 msec response time Calculated Volume Air Flow Displays in cubic feet per minute CFM or cubic meters per minute CMM Cross Sectional Area 20 Sq Inches default Programmable from 1 to 400 Sq Inches via keypad High Alarm Set Point Velocity Maximum range default Programmable via keypad Low Alarm Set Point Velocity Minimum Range default Programmable via keypad Hig...

Page 25: ...to 120m 400 LED Indicators Green TX LED Transmit Red RX LED Receive Data Transmission Interval 2 seconds default Settable from 1 to 60 seconds through End Device Configuration software Unit ID Address Default is 1 Settable from 1 to 48 through End Device Configuration software Wireless Receiver UWTC REC1 Power USB 5V Powered 300 mA max USB Compatibility USB 1 1 USB 2 0 LED Indicators TX transmit R...

Page 26: ...d to be operated in a clean and dry indoor environment Care should be taken to prevent the components of your wireless system from being exposed to moisture toxic chemicals and extreme cold or hot temperature that are outside the specifications listed in this manual 7 1 1 Operating Conditions The following is a list of basic good practice you should apply when operating your wireless system Never ...

Page 27: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Page 28: ...ors M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Communications Based Acquisition Systems M U Data Logging Systems M U Wireless Sensors Transmitters Receivers M U Signal Conditioners M U D...