INTRODUCTION

This  instrument  is  design  to  measure  the  differ-

ence between the pressures applied on the two ports 

of the meter. With a Pitot tube (separate purchasing) 

the meter also measures air speed.   

Besides displaying result of current measurement, 

and hold function the meter also store max / min / 

max-min / avg values. 

SAFETY INFORMATION 

Only  operate the  measuring instrument  properly, 

for its intended purpose and within the units speci-

fied in the technical data. Do not use force.   

Never  store  the  product  together  with  solvents, 

acids or other aggressive substances.   

Only carry out the maintenance and repair work 

that is described in the documentation. Follow the 

prescribed steps when doing so. 

SPECIFICATIONS 

Unit

 

Range

 

Resolution

 

mmH

2

O

 

0 to 1019.7 

0.1

 

mmHg

 

0 to 75.00 

0.01

 

psi

 

0 to 1.450 

0.001

 

inH

2

O

 

0 to 40.14 

0.01

 

inHg

 

0 to 2.953 

0.001

 

hPa

 

0 to 100.00 

0.01

 

mbar

 

0 to 100.00 

0.01

 

Pa

 

0 to 10000 

1

 

fpm

 

1

 

m/s

 

0.1

 

Accuracy: 

Nominal temperature 22°C, ±1dgt)

±0.03 hPa(0 to 0.30 hPa), 

±0.05 hPa(0.31 to 1.00 hPa), 

±(0.1 hPa+1.5% of reading) (rest of range) 

±0.3 mmH

2

O(0 to 3.0 mmH

2

O), 

±0.5 mmH

2

O(3.1 to 10.2 mmH

2

O), 

±(1.02 mmH

2

O + 1.5% of reading) (rest of range) 

Measurement rate: 

2 times/second.

Auto power off: 

10 minutes.

Low  battery  indication: 

The  “

”  is  displayed

when the battery voltage drops below the operat-

ing level. 

Operating  Environment: 

0°C  to  50°C  (32°F  to

122°F) at < 70% R.H. 

Storage  Temperature: 

-20°C  to  60°C  (-4°F  to

140°F), 0 to 80% R.H. with battery removed from 

meter. 

Battery:

 1.5Vx2pcs AAA size.

Battery Life:

85 hours typical.

Dimensions: 

24.7mm(T)  x  50.9mm(W)  x  132.9

mm(H). 

Weight:

Approx. 113g. (Including batteries)

OPERATIONS 

There  are  3  operation  modes,  namely  Measure-

ment Mode, Set Mode and Hold Mode. 

1. Turning on and off the meter:

When power is off, a short push on “ ” key turns

on the meter and enter Measurement Mode.

When power is on press and hold “ ” key for 2

seconds turns the power off.

Right  after  power  on,  all  the  indicators  on  the

LCD display lights up for one second.

2. Measurement Mode:

Before start measuring, disconnect the meter from

pressure sources, press “▲” key to zero the meter.

(When there is a measurement > 10hpa, Zeroing

function is disabled).

Connect  the  meter  to  the  pressure  sources,  the 

meter  start  measuring  and refresh  every 0.5  sec-

onds. 

Backlight:

In measurement mode, a short push on “ ” key 

turns  on  the  backlight  for  dark  environment,  it 

goes off automatically after 15 seconds if without 

further operation. 

3. Set Mode:

In set mode, user can select Unit , Density of Me-

dium and turn on/off APO function.(APO = Auto

Power Off, when APO is ON, the meter power off

automatically if no operation in 10 minutes.)

When power is off, press and hold “ ” key for 2

seconds to enter Set Mode.

3.1  Unit  setting:  Press  “▲”  key  to  change  unit

(press  and  hold  “▲”  key,  to  scroll  units), 

press “MODE” to selection the unit and enter 

next setting. If unit of air speed (fpm or m/s) 

is selected, then next setting is density setting 

otherwise APO setting. 

3.2 Density of the medium setting: Press “▲” key 

to  change  value  (press  and  hold  “▲”  key  to 

scroll  values),  press  “MODE”  key  to  select 

and enter APO On/Off setting (Max. Density: 

0.624lb/ft

3

 or 9.999kg/m

3

3.3  APO  ON/OFF:  short  push  on  “▲”  key 

switches 

between  ON  and  OFF.  Push  “MODE”  key  to 

select,  exit  Set  Mode  and  Enter  Measurement 

Mode.   

(Please  refer  to  the  section  of  Measurement 

Mode.) 

Note:

 To turn off power in Set Mode abandon the

current  setting  and  previous  settings  remain  un-

changed. If “

” indicator appeared, the setting

value can works till power off but did not save to

meter.

4. Hold Mode:

In Measurement Mode, short push “Mode” key to

enter  Hold  Mode  with  indicator  “HOLD”  shows

at upper left corner of LCD.

Right after entering this mode, shown in the LCD

is  the last  measurement and  the  reading  will  not

be updated with new measurement.

Push “Mode” key each time the display changes

in following sequence:

4.1 Last Measurement: with indicator “HOLD”.

4.2  Maximum  value  recorded:  with  indicator

“HOLD” + “MAX”. 

4.3  Minimum  value  recorded:  with  indicator 

“HOLD” + “MIN”. 

4.4 Maximum-Minimum value recorded: with in-

dicator “HOLD” + “MAX-MIN”. 

4.5  Average  of  values  recorded:  with  indicator 

“HOLD” + “AVG”. (only for wind speed) 

4.6  Exit  the  Hold  Mode  and  return  to  Measure-

ment Mode. 

To clear the recorded values: 

In  Hold  Mode  and  during  viewing  MAX,  MIN, 

MAX-MIN or AVG, press and hold “Mode” key 

for 2 seconds, to clear the recorded data and re-

turn to measurement mode. 

Installing and Replacing Battery 

A. Battery Cover

B. Battery

1. Power is supplied by 2pcs 1.5V (AAA SIZE).

2. The  “

”  appears  in  the  display  when  battery

replacement is needed.

3. Push the Battery Cover and lift it in the direction

as shown in the figure.

4. Remove the batteries from battery compartment.

5. Replace with 2 new AAA batteries with polarity

as  indicated  on  the  bottom  of  Battery  Compart-

ment.

6. Replace the Battery Cover.

Note:

 When not use for long time remove battery

and don’t keep in place with high Temp, or 

high humidity. 

Cleaning 

Periodically wipe the case with a damp cloth and 

detergent, do not use abrasives or solvents.   

HHC280

Differential Manometer

INSTRUCTION

SHEET

M5515/1015