background image

HH11C 

DIGITAL THERMOMETER

Servicing North America: 

U.S.A. Omega Engineering, Inc. 

Headquarters: Toll-Free: 1-800-826-6342 (USA & Canada only) 

Customer Service: 1-800-622-2378 (USA & Canada only) 

Engineering Service: 1-800-872-9436 (USA & Canada only) 

Tel: (203) 359-1660 Fax: (203) 359-7700 

e-mail: [email protected]

For Other Locations Visit omega.com/worldwide

It is the policy of OMEGA Engineering, Inc. to comply with all 
worldwide safety and EMC/EMI regulations that apply. OMEGA is 

constantly pursuing certification of its products to the 
European New Approach Directives. OMEGA will add the CE 
mark to every appropriate device upon certification. 
The information contained in this document is believed to be 
correct, but OMEGA accepts no liability for any errors it 
contains, and reserves the right to alter specifications without 
notice. 

WARNING: These products are not designed for use in, and 
should not be used for, human applications. 

INTRODUCTION

This  instrument  is  a  portable  4½  digit,  compact-sized 

digital  thermometer  designed  to  use  external  K-type  

thermocouples  as  temperature  sensor.  Temperature 

indication  follows  National  Bureau  of  Standards  and 

IEC  584  tem-perature/voltage  tables  for  K-type 

thermocouples.  One  K-type  thermocouple  is  supplied 

with the thermometer.

SAFETY INFORMATION 

It is recommended that you read the safety and  

operation instructions 

before using this t

hermometer. 

WARNING 

To avoid electrical shock, do not use this 

instrument when working voltages at the 

measurement surface over 24V AC or 60V DC.

WARNING 

To avoid damage or burns, do not make temperature 

measurement in microwave ovens.

CAUTION 

Repeated sharp flexing can break the thermocouple 

leads. To prolong lead life, avoid sharp bends in the 

leads, especially near the connector.

The 

  symbol  on  the  instrument  indicates  that  the 

operator must refer to an explanation in this manual. 

SPECIFICATIONS 

ELECTRICAL

Temperature Scale:

 Celsius or Fahrenheit user-selectable 

Measurement Range: 

-200°C to 1372°C, -328°F to 2501°F

Auto range: 

0.1°C/1°C, 0.1°F/1°F 

Accuracy:

Accuracy  is  specified  for  operating  tempera-

tures over the range of 18°C to 28°C (64°F to 82°F), for 

1 year, not including thermocouple error. 

±(0.1%rdg+0.5°C) on -60°C to 1372°C 

±(0.1%rdg+2°C) on -60°C to -200°C 

±(0.1%rdg+1°F) on -76°F to 2501°F 

±(0.1%rdg+4°F) on -76°F to -328°F 

Temperature Coefficient: 

0.1  times  the  applicable  accuracy  specification  per  °C 

from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 

82°F to122°F). 

Input Protection: 

60V dc or 24V ac rms maximum input voltage on any 

combination of input pins. 

Input Connector:   

Accepts  standard  miniature  thermocouple  connectors 

(flat blades spaced 7.9mm, center to center). 

ENVIRONMENTAL 

Ambient Operating Ranges: 

0°C to 50°C (32°F to 122°F) 

Storage Temperature: 

-20°C to 60°C (-4°F to 140°F) 

GENERAL 

Display:

 4½ digit liquid crystal display (LCD) with max-

imum reading of 19999. 

Low battery indication:

 The “

 

” is displayed when the 

battery voltage drops below the operating level. 

Measurement rate:

 1 time/second. 

Operating environment:

 0°C to 50°C at<70% R.H. 

Storage temperature:

 -20°C to 60°C, 0 to 80% R.H. with 

battery removed from meter. 

Accuracy:

 Stated accuracy at 23°C±5°C, <75% R.H. 

Battery:

 Standard 9V battery (NEDA 1604, IEC 6F22). 

Battery Life:

 100 hours typical with carbon zinc battery. 

Dimensions:

 210mm(H) x 65mm(W) x 35mm(D). 

Weight:

 approx. 282g including battery. 

Supplied Wire: 

4 feet type “K” thermocouple bead wire 

(

PTFE

  tape  insulated).  Maximum  insulation  tempera-

ture  260°C  (500°F).  Wire  accuracy  ±2.2°C  or  ±0.75% 

of reading (whichever is greater) from 0°C to 800°C. 

OPERATING INSTRUCTIONS

1.

“°C/°F” Button: Selecting the Temperature Scale

Readings are displayed in either degrees Celsius(°C)

 

or 

degrees Fahrenheit(°F).  When the thermometer is

turned on, it is set to the temperature scale that was in 

use when the  thermometer was last turned off. To

change the temperature scale, hold down for 2 seconds 

“APO >2S°C/°F” key.

2.

“HOLD” Button

Press the “HOLD” key to enter the Data Hold mode,

 

the 

“HOLD”  annunciator  is  displayed.  When  HOLD

mode is  selected,  the thermometer  held the present

readings and stops all further measurements. Press  the 

“HOLD” key again cancels HOLD mode, causing   the 

thermometer to resume taking measurements.

3.

“REL” Button

Press “REL”  key to enter Relative  mode,  zero the

display,  and store  the displayed  reading  as  a  reference 

value and annunciator  REL  is  displayed.  Pressing 

“REL” key over 2 seconds to exit the relative mode.

4.

“APO” Button

Pressing “APO >2S °C/°F” key to trigger  on or  off

APO  mode,  and then APO  annunciator  will appear  or 

disappear on the display. Power is automatically turn off, 

if no operation for a period of time, and “APO”

annunciator is displayed at upper-left corner when APO

function is enabled.

5.

“MIN/MAX” Button

Press “MIN/MAX” once to begin recording MIN and

MAX.  Press  “MIN/MAX”  to  select  MIN  or  MAX  or 

MAX-MIN  or  AVG.  Hold down for  2  seconds to exit 

MIN/MAX function.

In  the  MIN/MAX  record  mode  can  not  power  off, 

must leave MIN/MAX record mode then power off.