background image

 

 

 

www

e-mail: [email protected]

For latest product manuals:

.omegamanual.info

NORWALK, CT

Shop online at

omega.com

SM

User’s Guide

DPG4000

Digital Test Gauge

Summary of Contents for DPG4000

Page 1: ...www e mail info omega com For latest product manuals omegamanual info NORWALK CT Shop online at omega comSM User sGuide DPG4000 Digital Test Gauge ...

Page 2: ...specifications without notice omega com info omega com Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide ...

Page 3: ... protected to prevent unauthorized changes to co guration 1 1 Customer Service OMEGA Engineering Tel 203 359 1660 Fax 203 359 7811 www omega com email cservice omega com 1 2 Standard Equipment Check to see that your DPG4000 has arrived intact Batteries are factory installed unless you have purchased the optional 24V powered version in which case batteries are not supplied or installed Save the pac...

Page 4: ...irective 2002 96 EC Hazard Location Information Ex Hazardous Areas An Ex hazardous area as used in this manual refers to an area made hazardous by the potential presence of flammable or explosive vapors These areas are also referred to as hazardous locations see NFPA 70 Article 500 Misuse Should the DPG4000 be exposed to overpressure or sudden physical shock i e being dropped it should be examined...

Page 5: ...nstalled Connect a 24V power supply to the terminal block on the rear of the gauge noting proper polarity Do not install batteries when external power will be used Push the power button momentarily to turn the unit on Push it again to turn it off Set up and configuration Push the CONFIG button to access the user settable functions on the gauge Each time the CONFIG button is pressed the display adv...

Page 6: ...ttery voltage Actual voltage and a percent of life bargraph indicate battery condition 4 Display actual temperature The DPG4000 is temperature compensated this displays the temperature measured by the internal sensor The value can be set to degrees F or degrees C using the arrow keys 5 Set damping Choices are on and off set with the and keys Turning damping on will smooth readings from pulsating p...

Page 7: ...t LOC lock for those parameters you do not want to be accessible and UnLOC unlock for those can be accessed 3 In order the functions that can be unlocked locked or accessed are Zero function enable disable Set pressure units enable disable Auto shutdown adjustment enable disable Damping settings enable disable Sample rate setting enable disable Tare setting enable disable Custom engineering units ...

Page 8: ... seconds until CLr is displayed The analog bar graph at the bottom of the display indicates the applied pressure level relative to the full range of the gauge Keep in mind that if a TARE value has been programmed into the gauge the displayed pressure will not reflect the true pressure applied Changing the Batteries WARNING Explosion hazard Batteries must only be changed in an area known to be non ...

Page 9: ...or configuration calibration and to transfer measurement data from the gauge For detailed specifications on the interface and software communication see pages 14 and 15 in this manual WARNING The RS 232 interface must not be used in hazardous areas CLEANING To clean the DPG4000 use a cloth with a mild cleaning solution Specifications All specifications cover the temperature range from 0 C to 50 C ...

Page 10: ...ds and gases compatible with 316 stainless steel 1 PSI and compound ranges clean dry non corrosive gas Environmental Operating Temperature 10 C to 55 C Storage 20 C to 70 C 4 F to 158 F Humidity 10 to 95 RH Non condensing Pollution Degree II Mechanical Dimensions 4 5 diameter x 2 2 depth x 5 height Pressure Connection NPT Male Housing Stainless steel meets NEMA 4 IP65 Display 5 1 2 Digits 0 65 16 ...

Page 11: ...measurement deviates by more than 75 of the specified accuracy or 0 04 Test Equipment Verification and calibration of the DPG4000 requires pressure and or vacuum standards able to produce and indicate pressures from vacuum to the full scale range of the unit under test In order to maintain the specified accuracy of the DPG4000 standards should have a TUR of 4 1 or better Connections The DPG4000 us...

Page 12: ...ould look as shown in the illustration that follows Use the Pressure Standard to output 437 50 psi 87 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the Pressure Standard to output 375 00 psi 75 After the output has stabilized press the ENTER key to continue A...

Page 13: ...r the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the Pressure Standard to output 125 00 psi 25 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen should look as show...

Page 14: ...elow Note Only some ranges use vacuum calibration If your gauge is not one than this step will be automatically skipped and calibration will be complete Use the Pressure Standard to output 12 00 psi After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the unit will reset and power up in normal mode PSI 12 00 ...

Page 15: ... 10 546 21 093 70 309 210 93 351 54 703 09 1406 2 2109 3 3515 4 7030 9 mH 2 O 20 C 0 704336 0 7043 10 565 21 130 70 434 211 30 352 17 704 34 1408 7 2113 0 3521 7 7043 4 inH 2 O 4 C 27 68067 27 681 415 21 830 42 2768 1 8304 2 13840 27681 55361 83042 inH 2 O 20 C 27 72977 27 730 415 95 831 89 2773 0 8318 9 13865 27730 55460 83189 inH 2 O 60 F 27 70759 27 708 415 61 831 23 2770 8 8312 3 13854 27708 5...

Page 16: ... TARE Tares the offset pressure of the reading on the calibrator TARE Returns the current tare value PRES_UNIT Returns the pressure unit for the upper display PRES_UNIT Sets the pressure unit for the display ZERO_MEAS Zeros pressure of the calibrator ZERO_MEAS Returns the current zero offset value MINMAX_RST Resets the minimum and maximum recorded values MIN Returns the minimum recorded value MAX ...

Page 17: ...in bars mBar Pressure in millibars Kg cm2 Pressure in kilograms per centimeter squared InH2O4C Pressure in inches of water at 4 C InH2O20C Pressure in inches of water at 20 C InH2O60F Pressure in inches of water at 60 F mH2O4C Pressure in meters of water at 4 C MH2O20C Pressure in meters of water at 20 C cmH2O4C Pressure in centimeters of water at 4 C cmH2O4C Pressure in centimeters of water at 20...

Page 18: ...eter value received 105 Entry is above the upper limit of the allowable range 106 Entry is below the lower limit of the allowable range 108 A required command parameter was missing 109 An invalid pressure unit was received 117 An unknown command was received 120 The serial input buffer overflowed 121 Too many entries in the command line 122 Pressure module not connected ...

Page 19: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Page 20: ...s Air Velocity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity Equipment DATA ACQUISITION Communications Based Acquisition Systems Data Logging Systems Wireless Sensors Transmitters Receivers Signal Conditioners Data Acquisition Softwa...

Reviews: