background image

Accuracy

Includes linearity, hysteresis, repeatability 
±0.25% of full scale ±1 least significant digit
Sensor hysteresis:  ±0.015% FS, included in accuracy 
Sensor repeatability:  ±0.01% FS, included in accuracy

Display

3 readings per second nominal display update rate 
4 digit LCD, 0.5" H and 5 character 0.25" H alphanumeric 
BL:  Red LED backlight activated for 1 minute via front button 

Batteries, Battery Life, Low Battery Indication

B:   2 AA alkaline, approx. 2000 hours 
BL:  2 AA alkaline, approx. 150 to 1500 hours depending on 
 

backlight usage 

Low battery symbol on display

Auto Shutoff

User selectable 1 minute to 8 hours or front button on/off

Controls & Functions

Three front buttons turn gauge on or off, zeros gauge refer-
ence gauges, and cycles through functions

Memory

Six max-reading memory, user settable to MEM 1, 2, 3, 4, 5, 
6, or aircraft tire NLG 1, NLG 2, MLG 1, MLG 2, MLG 3, MLG 4

Calibration

Pass code protected calibration 
Non-interactive zero, span, and linearity, ±10% of range

Connection and Material

1/4” NPT male fitting, 316L stainless steel
All wetted parts are 316L stainless steel

Housing Material

DPG2001B:  Extruded aluminum case, epoxy powder coated,
 

ABS/polycarbonate bezel, front and rear gaskets,  

 

polycarbonate label

DPG2004B:  NEMA 4X ABS/polycarbonate case, rear gasket,
 

polycarbonate label

Overpressure, Burst, Vacuum

3000 psig sensor:  5000 psig overpressure, 10,000 psi burst 
5000 psig sensor:  7500 psig overpressure, 10,000 psi burst 
All other sensors:  2 X sensor range overpressure 
All other sensors:  4 X sensor range burst
Vacuum service: ±15 psig, 15 psia, 15 psig, 30 psia, 100 
psig, 100 psia, 200 psig sensors. Vacuum on all others will 
permanently damage sensor 
112.5% full scale out-of-range display: 1– – – or 1 –.–.–.– 
Under-range display (non-vacuum sensors): –Err

Weight

 (approximate)

Gauge: 

9 ounces 

Shipping weight:  1 pound

Environmental Temperatures

Storage: 

–40 to 203°F  (–40 to 95°C)

Operating: 

–4 to 185°F  (–20 to 85°C)

Compensated:  32 to 158°F   (0 to 70°C)

Standard

NEMA 4X

psi

kPa

MPa

mbar

bar

atm kg/cm

2

g/cm

2

mmH

2

O cmH

2

O oz/in

2

ftH

2

O inH

2

O mmHg

torr

inHg

DPG2001B-15A

DPG2004B-15A

15.00 to 0 absolute

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-30A

DPG2004B-30A

30.00 to 0 absolute

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-100A

DPG2004B-100A

100.0 to 0 absolute

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-15VAC

DPG2004B-15VAC

0 to –15.00 vacuum

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-15V15G

DPG2004B-15V15G

–15.0 to 15.0

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-15V100G

DPG2004B-15V100G

–15.0 to 100.0

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-15V200G

DPG2004B-15V200G

–15.0 to 200.0

4

4

4

4

4

4

4

4

4

DPG2001B-03G

DPG2004B-03G

0 to 3.000

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-05G

DPG2004B-05G

0 to 5.000

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-15G

DPG2004B-15G

0 to 15.00

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-30G

DPG2004B-30G

0 to 30.00

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-60G

DPG2004B-60G

0 to 60.00

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-100G

DPG2004B-100G

0 to 100.0

4

4

4

4

4

4

4

4

4

4

4

4

4

4

DPG2001B-200G

DPG2004B-200G

0 to 200.0

4

4

4

4

4

4

4

4

4

DPG2001B-300G

DPG2004B-300G

0 to 300.0

4

4

4

4

4

4

4

4

DPG2001B-500G

DPG2004B-500G

0 to 500.0

4

4

4

4

4

4

4

DPG2001B-1000G

DPG2004B-1000G

0 to 1000

4

4

4

4

4

4

4

DPG2001B-2000G

DPG2004B-2000G

0 to 2000

 

4

4

4

4

4

4

DPG2001B-3000G

DPG2004B-3000G

0 to 3000

4

4

4

4

4

4

DPG2001B-5000G

DPG2004B-5000G

0 to 5000

4

4

4

4

Options for all models

Add to end of model number

-BL

Backlighting (red LED)

-NIST

5-point NIST calibration

2.88"

2.0"

1.65"

NEMA 4X

3.0"

Standard

¼" NPT

3.38" Standard 

3.5" NEMA 4X

0.75"

X X X X X

t

s

Zero/Clear

Memory

X X X X X

88888

2

Up 

Zero/Clear

Numeric 

display

Lower 

alpha- 

numeric 

display

Down 

Memory

Power

Low Battery 

Symbol

Minus Sign

DPG2004B, NEMA 4X 

DPG2004BBL, NEMA 4X

DPG2001B 

DPG2001BBL

DPG2001B, DPG2004B  Series

INSTRUCTION

SHEET

Shop online at omega.com  e-mail: [email protected] 

For latest product manuals: www.omegamanual.info

DPG2001B 

Battery Powered, Min/Max/Zero 

DPG2001BBL  Battery Powered, Min/Max/Zero, Display Backlighting 

DPG2004B 

Battery Powered, Min/Max/Zero, NEMA 4X 

DPG2004BBL  Battery Powered, Min/Max/Zero, NEMA 4X, Display Backlighting

M4999/0818

Digital Pressure Gauges

Ranges and Resolution 

Resolution is fixed as indicated in table. 

G

 

Gauge reference pressure

VAC

 

Vacuum gauge, minus sign not used

A

 

Absolute reference (normally reads atmospheric pressure, reads zero at full vacuum)

TM

Reviews: