background image

DP41-B

Isolated Analog Output Option

Servicing North America:

USA:

 

         One Omega Drive, P.O. Box 4047

ISO 9001 Certified

 

        Stamford CT 06907-0047
        TEL: (203) 359-1660 

FAX: (203) 359-7700

e-mail: [email protected]

Canada:

 

         976 Bergar

        Laval (Quebec) H7L 5A1
        TEL: (514) 856-6928 

FAX: (514) 856-6886

e-mail: [email protected]

For immediate technical or application assistance:

USA and Canada:

          Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA

®

        Customer Service: 1-800-622-2378 / 1-800-622-BEST

®

        Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN

®

        

TELEX: 996404   EASYLINK: 62968934   CABLE: OMEGA

Mexico and

  

         TEL: (001)800-TC-OMEGA

®

 

FAX: (001) 203-359-7807 

Latin American: 

        

En Español: (001) 203-359-7803 
e-mail: [email protected]

Servicing Europe:

Benelux:

  

Postbus 8034, 1180 LA Amstelveen, The Netherlands
TEL: +31 20 3472121 

 

FAX: +31 20 6434643

Toll Free in Benelux: 0800 0993344 
e-mail: [email protected]

Czech Republic:

 

        Rudé armády 1868, 733 01 Karviná 8
        TEL: +420 69 6311899 

FAX: +420 69 6311114

e-mail: [email protected]

France:

 

         11, rue Jacques Cartier, 78280 Guyancourt

        TEL: +33 1 61 37 29 00  FAX: +33 1 30 57 54 27
        Toll Free in France: 0800 466 342

e-mail: [email protected]

Germany/Austria:

          Daimlerstrasse 26, D-75392 Deckenpfronn, Germany

        TEL: +49 7056 9398-0 

FAX: +49 7056 9398-29

        Toll Free in Germany: 0800 639 7678

e-mail: [email protected]

United Kingdom:

 

        One Omega Drive

ISO 9002 Certified

 

        River Bend Technology Centre
        Northbank, Irlam  Manchester M44 5BD United Kingdom  
        TEL: +44 161 777 6611  FAX: +44 161 777 6622
        Toll Free in England: 0800 488 488

e-mail: [email protected]

OMEGAnet

®

 On-Line Service

www.omega.com

Internet e-mail

[email protected]

®

®

M2548/0

8

02

WARRANTY/DISCLAIMER

OMEGA ENGINEERING,  INC.  warrants  this  unit  to  be  free  of  defects  in
materials  and  workmanship  for  a  period  of 

one (1)  year

from  the  date  of

purchase.  In  addition  to  OMEGA’s  standard  warranty  period,  OMEGA
Engineering will extend the warranty period for 

four (4) additional years

if the

warranty card enclosed with each instrument is returned to OMEGA.

If  the  unit  malfunctions,  it  must  be  returned  to  the  factory  for  evaluation.
OMEGA’s Customer Service Department will issue an Authorized Return (AR)
number  immediately  upon  phone  or  written  request.  Upon  examination  by
OMEGA, if the unit is found to be defective, it will be repaired or replaced at
no  charge.  OMEGA’s  WARRANTY does  not  apply  to  defects  resulting  from
any action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or unauthorized
modification. This WARRANTY is VOID if the unit shows evidence of having
been tampered with or shows evidence of having been damaged as a result of
excessive  corrosion;  or  current,  heat,  moisture  or  vibration;  improper
specification;  misapplication;  misuse  or  other  operating  conditions  outside  of
OMEGA’s  control.  Components  which  wear  are  not  warranted,  including  but
not limited to contact points, fuses, and triacs.

OMEGA is  pleased  to  offer  suggestions  on  the  use  of  its  various
products.  However,  OMEGA neither  assumes  responsibility  for  any
omissions  or  errors  nor  assumes  liability  for  any  damages  that  result
from the use of its products in accordance with information provided by
OMEGA,  either  verbal  or  written.  OMEGA warrants  only  that  the  parts
manufactured  by  it  will  be  as  specified  and  free  of  defects.  OMEGA
MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND
WHATSOEVER,  EXPRESS  OR  IMPLIED,  EXCEPT  THAT  OF  TITLE,  AND
ALL

IMPLIED  WARRANTIES  INCLUDING  ANY

WARRANTY

OF

MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE
HEREBY DISCLAIMED.  LIMITATION  OF  LIABILITY:  The  remedies  of
purchaser set forth herein are exclusive, and the total liability of OMEGA
with  respect  to  this  order,  whether  based  on  contract,  warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed
the purchase price of the component upon which liability is based. In no
event  shall  OMEGA be  liable  for  consequential,  incidental  or  special
damages.

CONDITIONS:    Equipment  sold  by  OMEGA is  not  intended  to  be  used,  nor
shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in
or with any nuclear installation or activity; or (2) in medical applications or used
on humans. Should any Product(s) be used in or with any nuclear installation
or  activity,  medical  application,  used  on  humans,  or  misused  in  any  way,
OMEGA

assumes  no  responsibility  as  set  forth  in  our  basic

WARRANTY/DISCLAIMER  language,  and,  additionally,  purchaser  will
indemnify  OMEGA and  hold  OMEGA harmless  from  any  liability  or  damage
whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS / INQUIRIES

Direct all warranty and repair requests/inquiries to the OMEGA Customer Service
Department.    BEFORE  RETURNING  ANY PRODUCT(S)  TO  OMEGA,
PURCHASER  MUST OBTAIN  AN  AUTHORIZED  RETURN  (AR)  NUMBER
FROM  OMEGA’S  CUSTOMER  SERVICE  DEPARTMENT (IN  ORDER  TO
AVOID  PROCESSING  DELAYS).  The  assigned  AR  number  should  then  be
marked on the outside of the return package and on any correspondence.
The  purchaser  is  responsible  for  shipping  charges,  freight,  insurance  and
proper packaging to prevent breakage in transit.

FOR 

WARRANTY

RETURNS, please

have the following information
available BEFORE 
contacting OMEGA:

1. P.O. number under which the

product was PURCHASED,

2. Model and serial number of the

product under warranty, and

3. Repair instructions and/or specific

problems relative to the product.

FOR 

NON-WARRANTY

REPAIRS,

consult OMEGA for current repair
charges.  Have the following
information available BEFORE
contacting OMEGA:

1. P.O. number to cover the COST of

the repair,

2. Model and serial number of product,

and

3. Repair instructions and/or specific

problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever
an  improvement  is    possible.  This  affords  our  customers  the  latest  in
technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
©  Copyright  2002  OMEGA ENGINEERING,  INC.  All  rights  reserved.    This
document may not be copied, photocopied, reproduced, translated, or reduced
to  any  electronic  medium  or  machine-readable  form,  in  whole  or  in  part,
without prior written consent of OMEGA ENGINEERING, INC.

PATENT  AND  TRADEMARK  NOTICE:

This  product  is  covered  by  one  or

more  of  the  following  patents:    U.S.  Pat.  No.  Des.  336,895;  5,274,577;
6,243,021  /  CANADA 2052599;  2052600  /  ITALY 1249456;  1250938  /
GERMANY DE 41 34398 C2 / SPAIN 2039150; 2048066 / UK Patent No. GB2
249 837; GB2 248 954 / FRANCE BREVET NO. 91 12756. The “Meter Bezel
Design” is a trademark of Newport Electronics, Inc. USED UNDER LICENSE.
Other U.S. and International Patents pending or applied for.

USA

MADE

IN

5

6

SETTING OUTPUT SCALE AND OFFSET 

(continued)

Out 1

(OUTPUT 1)

Press ‘

RESET/ENTER

14) Display flashes 1st digit of
previous "

OUt 1

" value.

Press ‘

/MAX

’ & ‘ 

/MIN

15) Enter "

OUt 1

" value.

(Example 0000.00).

Press ‘

RESET/ENTER

16) Display advances to
"

REAd 2

" Reading 2 menu.

REAd 2

(READING 2)

Press ‘

RESET/ENTER

17) Display flashes 1st digit
of previous "

Reading 2

" value.

Press ‘

/MAX

’ & ‘ 

/MIN

18) Enter "

Reading 2

” value.

(Example 999999).

Press ‘

RESET/ENTER

19) Display advances to 
"

OUt 2

" Output 2 menu.

Out 2

(OUTPUT 2)

Press ‘

RESET/ENTER

20) Display flashes 1st digit of
previous "

OUt 2

" value.

Press ‘

/MAX

’ & ‘ 

/MIN

21) Enter "

OUt 2

" value.

(Example 0010.00).

Press ‘

RESET/ENTER

22) Display Display shows
"

StOREd

" stored message

momentarily and then advances
to the "

SP CNF

" SetPoint

Configuration menu.

The above example is for 0-10 V of the entire
range of the process input and analog output. For
0-20mA output you need to set "

MOdE

" to "

CURRNt

"

DC-Current output and "

OUt 2

" to 0020.00.

SPECIFICATIONS

Analog Output:

0-5 V, 1-5 V, 0-10 V, 0-20 mA, 4-20 mA level; 14-bit
resolution; 0.1% accuracy; 6 msec step response.

Isolation

:

Power to Input / Analog, Comm, Ethernet, & Relay Output:
2500 Vac per 1 min. test
Relays to Inputs / Analog, Comm, & Ethernet Outputs:
2500 Vac per 1 min. test
Ethernet to Inputs / Analog Outputs:
1500 Vac per 1 min. test
Between Inputs / Analog & Comm Outputs:
500 Vac per 1 min. test

OPERATION MANUAL

CONFIGURING THE OUTPUT 

(continued)

MOdE

ENTER OUTPUT MODE CONFIGURATION SUBMENU:

Press ‘

RESET/ENTER

5) Display shows " 

StOREd

"

stored message momentarily
and advances to "

MOdE

" to

allow users to select Output
type as Voltage (0-10Vdc or
Current (4-20mA-dc) source.

Press ‘

RESET/ENTER

6) Display flashes "

VOLt

"

(default) DC-Voltage output
option or previous selection
("

CURRNt

" DC-Current output).

Out.tyP

ENTER OUTPUT TYPE CONFIGURATION SUBMENU:

Press ‘

RESET/ENTER

7) Display shows "

OUt.tyP

"

Output Type Submenu.

Press ‘

RESET/ENTER

8) Display flashes "

UNF ILt

"

Unfiltered option or previous
selection.

Press ‘

/MAX

9) Scroll though and select
the following options for your
choice of setting:

"

UNF Ilt

": Transmits the unfiltered value of your signal input.

"

FILtRd

" : Transmits the filtered value of your signal input.

"

PEAk

" : Transmits the recorded Peak "

HI RdG

" Reading

value(s).

or "

VALLy

" : Transmits the recorded Valley"

LO RdG

"

Reading value(s).

SETTING OUTPUT SCALE AND OFFSET

OT.SC.OF

ENTER OUTPUT SCALE & OFFSET

CONFIGURATION SUBMENU:

REAd 1

(READING 1)

Press ‘

RESET/ENTER

10) Display shows "

REAd 1

"

Reading 1 menu.

Press ‘

RESET/ENTER

11) Display flashes 1st digit of
previous "

Reading 1

" value.

Press ‘

/MAX

’ & ‘ 

/MIN

12) Enter "

Reading 1

" value.

(Example 000000).

Press ‘

RESET/ENTER

13) Display advances to
"

OUt 1

" Output 1 menu.

Ot.SC.OF

(Output Scale & Offset)

REAd  1

value

OUt  1

value

REAd  2

value

OUt  2

value

FILtRd

(Filtered)

OUt.tyP

(Output Type)

UNFILt

(Unfiltered)

PEAk

VALLy

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations
that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach  Directives.  OMEGA will  add  the  CE  mark  to  every  appropriate  device  upon
certification.

This device is marked with the international hazard symbol. It is important to read
the  Setup  Guide  before  installing  or  commissioning  this  device  as  it  contains
important information relating to safety and EMC.

WARNING: 

These  products  are  not  designed  for  use  in,  and  should  not  be  used 

for, patient connected applications.