User’s Guide

CL-355A

Dry-Block Calibrator

Shop online at

®

MADE IN UK

Summary of Contents for CL-355A

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info User sGuide CL 355A Dry Block Calibrator Shop online at MADE IN UK ...

Page 2: ......

Page 3: ...handle even at maximum operating temperatures Omega s calibrator has been designed to comply with all relevant Radio Frequency interference and electrical safety regulations Specification Figures quoted are at the base of the well at the time of calibration Temperature range 5 C 9 F above ambient to 400 C 752 F Over temperature limit 450 C 842 F approximately Display resolution 0 1 Accuracy 0 6 C ...

Page 4: ...ovided by the equipment to the operator may be impaired All Omega Engineering units have been designed to conform to international safety requirements and are fitted with a preset over temperature cutout If a safety problem is encountered switch off at the mains socket and remove the plug from the supply 3 Installation 1 All Omega units are supplied with a power cable 2 Before connecting the mains...

Page 5: ...ovided by the equipment to the operator may be impaired All Omega Engineering units have been designed to conform to international safety requirements and are fitted with a preset over temperature cutout If a safety problem is encountered switch off at the mains socket and remove the plug from the supply 3 Installation 1 All Omega units are supplied with a power cable 2 Before connecting the mains...

Page 6: ...hes the set point allow the block to stabilize for 15 minutes before performing a calibration 4 The controller is factory pre set and only the set temperature should be changed THE CONTROLLER MUST NOT BE SET TO CONTROL ABOVE 400 C 752 F OR DAMAGE TO THE UNIT MAY RESULT After use 1 When you have finished calibrating remember that parts of the unit and the probe may be very hot so precautions should...

Page 7: ......

Page 8: ...he block to stabilize for 15 minutes before performing a calibration 4 The controller is factory pre set and only the set temperature should be changed THE CONTROLLER MUST NOT BE SET TO CONTROL ABOVE 430 C 806 F OR DAMAGE TO THE UNIT MAY RESULT After use 1 When you have finished calibrating remember that parts of the unit and the probe may be very hot so precautions should be taken to prevent inju...

Page 9: ...uitably qualified personnel If the fuses blow persistently a serious fault is indicated and you may need to return the unit to your supplier for repair ADDITIONAL INFORMATION The controller is factory preset See the Controller book for further details For safety reasons the configuration settings of the controller are password protected Please contact your dealer for further information 7 Replacem...

Page 10: ... to return the unit to your supplier for repair ADDITIONAL INFORMATION The controller is factory preset See the Controller book for further details For safety reasons the configuration settings of the controller are password protected Please contact your dealer for further information Listed below are important control parameters for the unit Level 1 band 11 int t 1 5 dEr t 12 DAC 1 0 Cyc t 0 3 Bn...

Page 11: ...ive it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a...