background image

 

 

www.microscopenet.com

 

 

2.5  Replacing the fuse 

1)  Turn off the power switch and disconnect the power cord. 
2)  Find the fuse holder at the rear bottom of the microscope body. 
3)  Pry out the fuse holder gently with a flat head screwdriver. 
4)  Replace the old fuse with a new one, and then insert it back in. See 

Fig. 2

 
 
 
 
 
 
 
 

 
 

3

 

Operation 

3.1  Adjusting illumination 

1)  Plug the power cord into the power socket on the microscope and connect it to the 

power outlet. 

2)  Turn on the power switch.   
3)  Rotate the brightness intensity dial to increase or decrease the brightness of the 

illuminator. 

Caution:

   

A diffusion filter is attached beneath the condenser to get uniform light and protect 
your eyes from strong LED light when a low power objective applies. The diffusion 
filter can be swung out to make the view field brighter when observing with a high 
power objective, such as 100X objective. 

3.2  Placing specimen 

1)  Place the slide on the mechanical stage.   
2)  Use the slide holder to gently secure the slide.   
3)  Turn the X and Y stage moving knobs to position the specimen in the center of 

viewing field. 

Caution:

    

Be sure not to allow an objective to touch a specimen slide when changing objectives. 

3.3  Focusing 

1)  With the 10X objective in position, raise the mechanical stage using the coarse focus 

knob until the specimen is close to the objective.   

2)  Turn the coarse focus knob until the specimen is in focus. Then use the fine focus 

knob to obtain a sharp image. You may now switch to another magnification 
objective. 

3)  To get a good focused image, you may need to combine the focus knob adjustment 

and interpupillary distance adjustment, along with eyepiece diopter adjustment stated 
in 

3.4

 and 

3.5

.   

Fig.2 

Fuse Holder 

Fuses 

Summary of Contents for M8211E

Page 1: ...User Manual Binocular Compound LED Microscope Model M8211E MicroscopeNet com...

Page 2: ...Table of Contents I Caution 1 II Care and Maintenance 2 1 Components Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 7 5 Troubleshooting Guide 8...

Page 3: ...he container should be retained should the microscope ever requires reshipment 3 Keep the instrument out of direct sunlight high temperature or humidity and dusty environments Ensure that the microsco...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...er Switch 3 Eyepiece Tube 10 Nosepiece 17 Brightness Intensity Dial 4 Viewing Head 11 Objective 18 Condenser Lock Thumb Screw 5 Slide Holder 12 Microscope Body 19 Condenser Focus Knob 6 Mechanical Sta...

Page 6: ...at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the holes on the nosepiec...

Page 7: ...filter can be swung out to make the view field brighter when observing with a high power objective such as 100X objective 3 2 Placing specimen 1 Place the slide on the mechanical stage 2 Use the slide...

Page 8: ...Adjusting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objecti...

Page 9: ...uadruple Objectives Achromatic 4X 10X 20X spring 40X spring Condenser Abbe NA 1 25 w iris aperture diaphragm and filter holder Rack and pinion focus adjustment Focus Mechanism Coaxial coarse and fine...

Page 10: ...slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter a...

Reviews: