Chapter 7 Maintenance and Repair
FLUSHING PUMP OFP-2
48
3.
Turn the OFP-2 power supply ON and confirm that the control panel indicators illuminate and
the standby LED then illuminates green.
4.
Press the footswitch and confirm that the pump runs.
5.
Lift the pump head lever and confirm that the standby LED turns to orange and the pump
head does not rotate.
6.
Confirm that the flow rate is sufficient following the guidelines in Section “Checking the Flow
WARNING
If the events described in the above steps 3 to 6 are not observed,
immediately switch the unit's supply OFF, disconnect the power cord
and contact Olympus.
7.4 Fuse Replacement
WARNING
• The fuse must always be replaced with the fuse specified within
“Care and Storage” on page 44. Failure to do so may cause malfunction
or failure of the OFP-2, causing a fire or electric shock hazard.
• Turn the OFP-2 OFF and unplug the power cord from the back of the
unit prior to removing the fuse drawer.
1.
Turn the OFP-2 OFF and disconnect the power cord from the mains power inlet.
2.
Referring to Figure 7.4, press the tab downwards to release the fuse drawer, pull out the
drawer and replace both fuses.
Figure 7.3
Drive slot