![Olympus Magna-Mike 8600 User Manual Download Page 103](http://html1.mh-extra.com/html/olympus/magna-mike-8600/magna-mike-8600_user-manual_3943437103.webp)
DMTA-10026-01EN, Rev. D, January 2015
Using Special Functions
93
Figure 5-2 The DIFFERENTIAL SETUP screen
4.
Press
[MEAS]
to return to the measurement screen with the displayed differential
value.
5.2
Using the Minimum/Maximum Thickness Mode
The Minimum thickness mode of the Magna-Mike 8600 is often used for scanning for
the minimum thickness. While scanning, a false thickness reading can be displayed if
the probe is not properly aligned with the surface. Using the Minimum mode greatly
reduces false readings because, when the probe is properly aligned, it will always
display the minimum thickness.
You can activate the Minimum/Maximum (
MIN/MAX)
thickness mode to also
display retained minimum and/or maximum thickness values. The
MIN
and/or
MAX
values appear on the right-hand side of the main thickness reading (see Figure 5-3 on
page 94). The user can select which measurement is shown in large font (Live,
Minimum or Maximum).
Summary of Contents for Magna-Mike 8600
Page 7: ...DMTA 10026 01EN Rev D January 2015 Table of Contents vii List of Tables 169 Index 171...
Page 8: ...DMTA 10026 01EN Rev D January 2015 Table of Contents viii...
Page 10: ...DMTA 10026 01EN Rev D January 2015 List of Abbreviations x...
Page 28: ...DMTA 10026 01EN Rev D January 2015 Introduction 18...
Page 70: ...DMTA 10026 01EN Rev D January 2015 Chapter 3 60...
Page 112: ...DMTA 10026 01EN Rev D January 2015 Chapter 5 102...
Page 120: ...DMTA 10026 01EN Rev D January 2015 Chapter 6 110...
Page 146: ...DMTA 10026 01EN Rev D January 2015 Chapter 7 136...
Page 160: ...DMTA 10026 01EN Rev D January 2015 Chapter 8 150...
Page 164: ...DMTA 10026 01EN Rev D January 2015 Chapter 9 154...
Page 168: ...DMTA 10026 01EN Rev D January 2015 Appendix A 158...
Page 178: ...DMTA 10026 01EN Rev D January 2015 List of Figures 168...
Page 180: ...DMTA 10026 01EN Rev D January 2015 List of Tables 170 Table 31 Calibration kits 163...
Page 186: ...DMTA 10026 01EN Rev D January 2015 Index 176 WINXL about 137...