1.
From the setup menu, activate the gate (see “Activating the Gates” on page 71).
2.
Position the gate using
G1 Start
,
G1 Width
, and
G1 Level
Dialog Boxes and Setup Pages” on page 19 for details on using the gate dialog
box).
3.
Go to the sidebar menu and select
Gate1
or
Gate2
page 24 for details on using the sidebar menu).
4.
Select
Zoom
to zoom in.
5.
Select
Zoom
again to zoom out (default view).
7.9
Gate Alarms
The EPOCH 6LT features a variety of alarm configurations for each measurement
gate. In unrectified (
RF
) mode, gate alarms can be used in positive, negative, or dual
gate modes.
By default, when an alarm condition is triggered, the EPOCH 6LT emits an audible
beep. The instrument also illuminates the red indicator, above the display window,
corresponding to the gate with which the alarm has been triggered. To toggle the
audible alarm on and off, see “Instrument Setup Page” on page 50.
The three primary types of gate alarms are as follows:
•
Positive threshold
•
Negative threshold
•
Minimum depth
7.9.1
Threshold Alarms
Threshold alarms can be set on gate 1 or gate 2.
A positive logic alarm is triggered when a signal exceeds the gate threshold. A
negative logic alarm is triggered when a signal falls below the gate threshold.
When you set a threshold alarm, the appearance of the tick marks at the end of the
gate changes (see Figure 7-9 on page 83).
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...