Pulser/Receiver Adjustments
63
Figure 6-3 Pulser dialog box
6.2.1
Energy (Voltage)
The EPOCH 6LT can adjust the pulse energy from 0 V to 400 V in increments of 100 V.
You can set the pulse energy to a minimum to extend the battery life, or set very high
pulse energy for the most difficult materials.
To adjust the pulse energy
1.
In the pulser dialog box, press the Enter key to highlight the
Energy
value.
2.
Turn the adjustment knob to vary the value (see “Navigating Dialog Boxes and
Setup Pages” on page 19).
In the
Energy
adjustment, coarse and fine steps are equal (100 V).
To maximize instrument battery life and transducer life, use lower energy settings
when the application permits. For most applications, the energy setting does not need
to exceed 200 V.
NOTE
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...