DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
264
Figure 11-27 Storing the T
2
point
11. Press
Clear
(P5) to complete the calibration and move to the Inspect mode.
12. If you are not satisfied with the collected points, you can overwrite a particular
point using the previous parameter keys (P1, P2, or P3), or press
Clear
(P5) to
clear the entire calibration and begin again.
11.5.3.2
Crack Sizing
Once the API 5UE software has been properly calibrated to a reference notch, use the
following steps to size a crack using Manual mode.
To size a crack using Manual mode
1.
Find the signal from the potential defect bring to maximum amplitude (see API’s
Recommended Practice 5UE for crack scanning and inspection requirements).
2.
If necessary, use 2
ND
F, (AUTO XX%) to bring the peak amplitude to 80 % of FSH
to ensure the peak amplitude is displayed on the screen.
3.
Press
RefAMax
(P2) to assign this peak amplitude value as A
max
. Note the value
of the peak amplitude in %FSH.
Summary of Contents for EPOCH 650
Page 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Page 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Page 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Page 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Page 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Page 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Page 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Page 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Page 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Page 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Page 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Page 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...