DMTA-10055-01EN, Rev. A, February 2015
Software Features and Options
237
Figure 11-11 DAC curves with adjusted gain
11.2.3.3
Transfer Correction
Transfer Correction is an adjustment in the reference gain setting during calibration of
the instrument, and is typically added when the surface conditions between a
calibration block and test piece are different. The coupling conditions on the test
surface can often cause signal loss after calibrating a DAC curve, which results in
inaccurate comparisons of the test reflectors with the calibrated DAC curve. The
EPOCH 650 can be easily adjusted for this potential difference by adding transfer
correction to the calibrated base gain after completing the DAC curve setup.
To add transfer correction to a completed DAC curve
1.
Choose
Basic
.
2.
Press dB.
3.
Use the Up and Down arrow keys or the adjustment knob to bring the scanning
gain to the desired level for transfer correction.
4.
Press
Add
(P2 key) to add the scanning gain to the base gain and to apply the
transfer correction.
Summary of Contents for EPOCH 650
Page 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Page 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Page 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Page 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Page 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Page 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Page 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Page 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Page 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Page 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Page 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Page 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...