DMTA
‐
10007
‐
01EN
[U8778365],
Rev. B,
June
2010
Basic
Operation
35
•
Mode
:
pulse
modes
available:
P/E
(pulse
‐
echo),
Dual
,
and
Thru
(through
‐
transmission)
•
Pulser
:
Pulser
waveforms
available:
Spike
or
Square
(tunable
square
wave)
•
Freq
:
Pulse
frequency
(square
wave
pulse
width)
ranges
from
0.1 MHz
to
20.00 MHz
The
Spike
pulse
is
equivalent
to
a
20
‐
MHz
square
wave
pulse.
2.2.4
Receiver
The
EPOCH 600
standard
receiver
functions
are
accessible
by
selecting
the
Rcvr
menu.
Each
individual
receiver
parameter
is
displayed
above
the
parameter
keys,
and
can
be
adjusted
by
pressing
the
corresponding
[P
<n>
]
parameter
key.
The
following
receiver
parameters
are
available
for
live
adjustments
in
the
Rcvr
menu:
•
Filter
:
Receiver
filter
setting
•
Rect
:
Waveform
rectification
(
Full
[fullwave],
Half+
[halfwave
positive],
Half–
[halfwave
negative],
RF
[no
rectification])
•
Reject
:
Percent
reject
(from
0 %
to
80 %)
The
eight
filters
available
on
the
EPOCH 600
allow
broadband
or
narrowband
settings
to
match
the
requirements
of
a
given
application.
Each
is
a
fully
digital
filter
set
with
the
following
low
‐
pass
and
high
‐
pass
cutoffs:
•
0.2 MHz–10 MHz
•
2.0 MHz–21.5 MHz
•
8.0 MHz–26.5 MHz
•
0.5 MHz–4.0 MHz
•
0.2 MHz–1.2 MHz
•
1.5 MHz–8.5 MHz
•
5.0 MHz–15 MHz
•
DC–10 MHz
Summary of Contents for EPOCH 600
Page 1: ...DMTA 10007 01EN U8778365 Revision B June 2010 EPOCH 600 Basic Operation Manual ...
Page 6: ...DMTA 10007 01EN U8778365 Rev B June 2010 vi List of Abbreviations ...
Page 18: ...DMTA 10007 01EN U8778365 Rev B June 2010 12 Important Information Please Read Before Use ...
Page 32: ...DMTA 10007 01EN U8778365 Rev B June 2010 26 Chapter 1 ...
Page 56: ...DMTA 10007 01EN U8778365 Rev B June 2010 50 Appendix ...
Page 60: ...DMTA 10007 01EN U8778365 Rev B June 2010 54 List of Tables ...