146 Chapter 9
Due to the acoustic characteristics of dual transducers, nonlinearity in the
distance calibration occurs as the thickness of the material decreases. The point of
maximum sensitivity is determined by the roof angle of the particular dual
transducer. It is recommended that the distance calibration is carried out using a
step block that covers the range of interest. Be careful when interpreting thickness
measurement readings made outside the calibrated range. The EPOCH 1000
Series does not have v-path correction; therefore, there might be some
nonlinearity within the calibrated range depending on the minimum thickness
used in the calibration process.
The zero offset value of dual element transducers can vary significantly at extreme
temperatures. If the temperature changes more than a few degrees from the
temperature at which the zero offset value was established, recheck its value. If
thickness measurements are to be made over a wide temperature range, it is strongly
recommended to use Olympus dual transducers that are designed for high-
temperature applications and that have built-in delay lines with a stable sound
velocity that does not change significantly with temperature. Specific
recommendations are Olympus D790-SM and D791 dual element transducers.
To calibrate using a dual element transducer
1.
Follow the initial setup procedure outlined in section 9.1 on page 134.
2.
Connect the transducer to an appropriate cable, and then connect the cable to the
conventional transducer connectors at the top of the instrument.
3.
Select
1/5 > Pulser > Mode = Dual
.
4.
Press
[GAIN]
, and then adjust the value very high so that the leading edges of the
back-wall echoes appear as nearly vertical lines on the screen.
5.
To use the leading edge when making thickness measurements, set the
measurement gate(s) to the edge detection mode:
a)
Choose
2/5 > Gate Setup > Setup
to open the
Gates
setup page.
b)
In the setup page, highlight
Gate
<n>
Measurement Mode
and set its value to
Edge
.
c)
Press
[ESCAPE]
to exit the setup page.
6.
Select
1/5 > Auto CAL > CAL Mode = Thickness
.
Downloaded from
Summary of Contents for EPOCH 1000 Series
Page 30: ...910 269 EN Rev B June 2011 18 Preface Downloaded from ManualsNet com search engine ...
Page 50: ...910 269 EN Rev B June 2011 38 Chapter 1 Downloaded from ManualsNet com search engine ...
Page 60: ...910 269 EN Rev B June 2011 48 Chapter 2 Downloaded from ManualsNet com search engine ...
Page 100: ...910 269 EN Rev B June 2011 88 Chapter 3 Downloaded from ManualsNet com search engine ...
Page 186: ...910 269 EN Rev B June 2011 174 Chapter 9 Downloaded from ManualsNet com search engine ...
Page 206: ...910 269 EN Rev B June 2011 194 Chapter 10 Downloaded from ManualsNet com search engine ...
Page 258: ...910 269 EN Rev B June 2011 246 Chapter 12 Downloaded from ManualsNet com search engine ...
Page 280: ...910 269 EN Rev B June 2011 268 Chapter 14 Downloaded from ManualsNet com search engine ...
Page 286: ...910 269 EN Rev B June 2011 274 Chapter 16 Downloaded from ManualsNet com search engine ...
Page 318: ...910 269 EN Rev B June 2011 306 Chapter 17 Downloaded from ManualsNet com search engine ...
Page 322: ...910 269 EN Rev B June 2011 310 Chapter 18 Downloaded from ManualsNet com search engine ...
Page 332: ...910 269 EN Rev B June 2011 320 Chapter 19 Downloaded from ManualsNet com search engine ...
Page 350: ...910 269 EN Rev B June 2011 338 List of Figures Downloaded from ManualsNet com search engine ...