ENDOCAPSULE 10 SYSTEM
How to register patient information
Q&A
Enter required patient information. Minimum
requirements are indicated by red * marks
and red numbers.
Click “Save & New” to continue registering.
Click “Save & Close” to return to main screen.
Note: In order to send patient messages,
register from Patient Guidance tab.
What is the recommendation for patient clothing?
Wear thin clothing without any metal. Wearing a top or bottom with
buttons or zippers is not recommended to avoid un-preferable effects
for the antenna.
If the recorder is removed from the cradle accidentally during
initialization or downloading, will the data be affected?
If the process is canceled, place the recorder on the cradle again and
restart the workstation. The data transfer may be interrupted, but the
data will be protected.
How long does it take to recharge the battery?
It takes about two hours. The green light will tell you when the battery
has finished charging.
How to initialize the recorder
How can I recognize when the battery needs to be replaced?
A message “Need new battery soon” will be displayed on the recorder.
Be sure to replace it before the message “Replace Battery” pops up.
Why does the display on the recorder disappear?
The display light will turn off to save battery power after 30 seconds
without operation.
How do I clean the equipment?
The antenna unit holder and recorder holder can be machine-washed.
Be sure to let them dry naturally and completely. Other equipment
such as the antenna unit, recorder, cradle and capsule activator should
have their surfaces wiped immediately after use with 70% ethyl or
isopropyl alcohol.