Olympus ENDOCAPSULE 10 SYSTEM Quick Start Manual Download Page 1

2

Preparation and settings for capsule endoscopy

ENDOCAPSULE 10 SYSTEM

1

3

5

6

7

When removing capsule from the 
container, be careful NOT to touch 
the top cover of capsule. Otherwise, 
clear images may not be captured.

Important notice to patients

4

After the capsule ingestion

8

9

9. Press “MENU” and select “Real Time View”. 

Confirm that images from the body are 
received properly. If necessary, images can 
be playbacked or captured.

10. Lock the button operation. Go to “Patient 

Message” and press “MENU” for at least 2 
sec. When patient guidance is activated, the 
patient does not necessarily need to touch 
the recorder. Beep or vibration will stop after 
30 sec.

12

11. Unlock the recorder. Press “MENU” for 

more than 2 sec. Open “Real Time 
View” to confirm that the examination is 
complete.

12. Place recorder into the cradle.
13. A message is displayed on the screen. 

Click “OK” to start downloading image 
data. When completed, a check mark is 
displayed on the main screen.

13

10

After the examination

11

Note: Check that the battery is 
fully charged (green lamp is lit).

1. Place recorder into the cradle.
2. Login to workstation. Register 

patient information and initialize 
recorder. 

Refer to the back page

3. Insert the antenna sheet into the 

cover between the two plates. 
Orange colored lines printed on the 
surface of the antenna sheet 
should face front. The cable always 
stay on the patient's left side.

4. Fasten the antenna belt around the 

patient’s body. The alignment hole 
should fit the umbilical area. Wrap 
the cable band around the black 
belt. Make sure that the cable and 
belt are tightly fixed.

5. Attach the recorder holder on the 

left side of the patient’s body. 
Connect the recorder and the 
antenna unit. Slide the connection 
part until it clicks.

6. Press “POWER” for at least 1 sec. 

Patient ID and name is displayed 
on the upper left screen.

7. Insert the capsule activator into the 

capsule container. The LED should 
start blinking.

8. Press “MENU” and select “Real 

Time View”. Confirm that images 
are received properly.

This sheet is for quick reference only. Please refer to the instructions provided with the device for instructions in details.

Antenna sheet

Plates

Before the examination

Disclaimer:  Any content or information (“Content”) presented herein is illustrative in nature and does not guarantee or represent specific information, outcomes, or results.  Olympus Corporation, its subsidiaries, affiliates, directors, 

officers, employees, agents, and representatives (collectively “Olympus”) does not represent to or warrant the accuracy or applicability of the Content.  Under no circumstances shall Olympus be liable for any costs, expenses, losses, 

claims, liabilities, or other damages (whether direct, indirect, special, incidental, consequential, or otherwise) that may arise from, or be incurred in connection with, the Content or any use thereof.

Reviews: