OLYMPUS CIX100 Cleanliness Inspector System
6
Calibrating White Balance for a new sample type
Make sure to log into the CIX ASW software as administrator. Then you have full
access to the [System configuration] on the start page of your CIX ASW software.
Calibrate the stage limits as described in the CIX ASW software’s help. Insert the
new sample type with the new sample, which you activated in the [Hardware] >
[Device List] > [Device settings] dialog box as described in chapter "Adding new
sample types to your CIX100 system" on page 13. These may be
l
tape lift,
l
particle trap, or
l
covered membrane.
For a list of samples and applications, please refer to chapter "Available sample
holders and sample types" on page 5 of this manual.
1. Click the [Calibration] button on the start page of CIX ASW.
24