background image

Complete Teardown, Cleaning, and Reassembly of the Olympus BHSU/BHTU Reverse Nosepiece Assembly

 

 

Revision 3

       

Page 10 of 19 

Remove the Center Pivot Bearing Ball 

Use a tweezers or other suitable tool 

to remove the ¼” 

bearing ball from the center bore in the stationary base 
(see 

Figure 18

).

 

 

Figure 18 

 Remove the center-pivot bearing ball 

Remove the Mechanical Detent Stop 

The next step is to remove the mechanical detent stop.  
Use a suitable JIS screwdriver to remove the two M2x4 
pan-head  screws  securing  this  stop  to  the  stationary 
base  of  the  turret  assembly  (see 

Figure  19

).    These 

screws  are  staked  in  place  with  adhesive  and  can  be 
surprisingly  stubborn,  so  be  sure  to  use  the  proper  JIS 
screwdriver  to  prevent  damaging  the  screw  heads.    It 
might also be helpful to heat the screws with a heat gun 
before  loosening  them,  but  do  not  melt  the  black 
knurled-rubber grip ring in the process. 

 

Figure 19 

 Remove the screws securing the detent stop 

Remove  the  loose  mechanical  detent  stop  from  the 
stationary base (see 

Figure 20

). 

 

Figure 20 

 Remove the mechanical detent stop 

Remove the Black Knurled-Rubber Grip Ring 

Carefully  remove  the  black  knurled-rubber  grip  ring 
from  the  outer  perimeter  of  the  revolving  turret  (see 

Figure 21

), being careful not to stretch or damage it in 

the process.  This ring is held onto the turret with a few 
spots of adhesive, so proceed carefully here.

 

 

Figure 21 

 Remove the knurled-rubber grip ring 

Remove the Threaded Retaining Ring 

The  stationary  base  is  held  inside  the  recess  of  the 
revolving  turret  by  the  threaded  retaining  ring  (see 

Figure 22

). 

 

Figure 22 

 Components of the turret assembly 

Summary of Contents for BHSU

Page 1: ...this PDF document in its entirety for personal non commercial purposes only provided that the contents are not modified in any way including the copyright notices contained herein This document may be...

Page 2: ...chanical Detent Stop 10 Remove the Black Knurled Rubber Grip Ring 10 Remove the Threaded Retaining Ring 10 Remove the Perimeter Bearing Balls 11 Remove the Stationary Base from the Turret 11 Thoroughl...

Page 3: ...n the slotted lock ring 9 Figure 16 Remove the slotted lock ring 9 Figure 17 Remove the pivot adjustment screw 9 Figure 18 Remove the center pivot bearing ball 10 Figure 19 Remove the screws securing...

Page 4: ...silicone RTV to revolving turret 15 Figure 45 Reinstall the knurled rubber grip ring 16 Figure 46 Remove any silicone RTV squeeze out 16 Figure 47 Place the wedge mount in position 16 Figure 48 Reinst...

Page 5: ...he Olympus BHSU and BHTU microscope stands Note that the original Olympus service literature did not address the teardown and repair of the various revolving nosepiece assemblies used on BH 2 stands a...

Page 6: ...st of solvents generally considered safe for nylon includes acetone diethyl ether heptane mineral spirits naphthalene and turpentine Before Starting with the Overhaul Before beginning the removal and...

Page 7: ...ing the wedge mount to the stationary base of the turret assembly see Figure 6 It is important to remove these four screws before attempting to remove the corrective optics assembly to prevent the bar...

Page 8: ...pivot adjustment screw unless you have access to such a tool since the slot in the lock ring will likely be damaged if you use an improper tool making removal of the lock ring much more difficult1 A...

Page 9: ...g the knurled rubber grip ring on the outer perimeter of the revolving turret and use the slotted screwdriver with the notched tip see Figure 13 to loosen the slotted lock ring securing the pivot adju...

Page 10: ...ent damaging the screw heads It might also be helpful to heat the screws with a heat gun before loosening them but do not melt the black knurled rubber grip ring in the process Figure 19 Remove the sc...

Page 11: ...th the threaded retaining ring do not fall out and become lost in the process see Figure 24 Figure 24 Remove the threaded retaining ring Remove the Perimeter Bearing Balls Carefully remove the 3 32 be...

Page 12: ...rease for the bearing balls Reinstall the Perimeter Bearing Balls Use tweezers to carefully set the 3 32 bearing balls there are 80 of them into the grease ring see Figure 29 placing the bearing balls...

Page 13: ...at the center pivot bore has been greased reinstall the bearing ball into the freshly greased center pivot bore The grease will hold the center pivot bearing ball in the proper position during subsequ...

Page 14: ...t with the other hand The motion of the turret should not feel gritty erratic or excessively stiff If it does loosen the slotted lock ring readjust the pivot adjustment screw and retighten the slotted...

Page 15: ...o the stationary base see Figure 43 Figure 43 Secure protective cover using three screws Reinstall the Knurled Rubber Grip Ring The final assembly step for the turret assembly is to reinstall the blac...

Page 16: ...sition While holding the wedge mount in this position place the threaded end of the corrective optics assembly through the center bore of the wedge mount and screw the corrective optics assembly into...

Page 17: ...anical detents thereby maximizing the useful service life of the reverse nosepiece assembly Follow the procedures detailed in this document to remove the reverse nosepiece assembly from the microscope...

Page 18: ...may be encountered if the threaded retaining ring has not been properly snugged down or if the center pivot adjustment screw has not been properly adjusted to remove the play in the center pivot mech...

Page 19: ...on 5 40 4 Bearing balls stainless steel 1 4 G25 25 count various Amazon 4 95 5 Pliers soft jaw Non Scratch Pliers Micro Mark 85161 34 95 Pliers soft jaw Tamiya 74061 Amazon 32 64 6 Lens spanner tool p...

Reviews: