
Chapter 7 Cleaning, Disinfection, and Sterilization Procedures
97
ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F
Figure 7.16
5.
Disconnect the ultrasonic cable from the endoscope connector.
6.
Disconnect the videoscope cable from the endoscope connector.
7.
Disconnect the endoscope connector from the light source.
8.
Transport the endoscope and the container holding the other items to the
reprocessing area.
7.4
Leakage testing
After precleaning, perform leakage testing on the endoscope to ensure that it is
waterproof.
Equipment needed
Prepare the following equipment, and wear appropriate personal protective
equipment.
• Personal protective equipment
• Large basin
• Clean water (see Section 6.2)
• Maintenance unit or light source
(MU-1 or CLV-180, CLV-160, CLV-U40)
• Leakage tester
(MB-155)
• Water-resistant cap (2 pcs)
(MH-553)