DMTA-10022-01EN, Rev. C, January 2015
Software Options
73
2.
In the
MEAS
screen, set the
MEAS MODE
to the desired echo detection mode:
STANDARD
,
AUTO E-TO-E
(optional), or
MANUAL E-TO-E
(only available
with the Waveform option).
When switching between standard and Echo-to-Echo modes, it may be necessary to
perform a second
[CAL ZERO],
because the zero offset is different for each
measurement mode. The 45MG can store two different zero calibrations, one for
standard mode, and the other for Echo-to-Echo mode.
3.
To perform the zero calibration again:
a)
Place a drop of couplant on the surface of the thin part of the test block.
b)
Couple the transducer to the thin part of the test block, and then press
[CAL ZERO]
.
c)
Once the thickness reading is stable, press
[ENTER]
.
d)
Using the arrow keys, edit the thickness value to match the known thickness
of the thin part of the test block.
6.2.1
Blanking Adjustments in Manual Echo-to-Echo Detection Mode
The 45MG offers two blanking functions to help detect valid echoes in situations
where material conditions generate unwanted signals:
EXT BLANK
The extended blank creates a blanked zone that begins at the left edge of the
waveform display, and in which no signals are detected. In situations where the
second or third pair of back-wall echoes are stronger or cleaner than the first pair,
use the extended blank to control which pair of echoes are used for measurement.
E1 BLANK
The echo 1 (E1) blank runs for a selected interval following the first detected echo.
Use the E1 blank to exclude any unwanted peaks occurring between the first and
second back-wall echoes. Unwanted peaks may be trailing edges of a large first
echo, or shear-wave reflections on thick test pieces. The E1 blank parameter is
only available in the manual Echo-to-Echo detection mode.
To adjust the extended and the E1 blank parameters
1.
Select the manual Echo-to-Echo mode:
NOTE
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...