DMTA-10022-01EN, Rev. C, January 2015
Maintaining and Troubleshooting the 45MG
201
INTERNAL MEMORY RESET
Deletes all stored data on the internal microSD memory card and reformats the
card.
The internal memory reset permanently deletes all thickness readings/waveforms that
are stored in the memory card. However, this reset does not delete any stored
transducer setups.
CLOCK RESET
Resets the date to 01/01/2010 in the MM/DD/YYYY format, and the time to 12:00
A.M. in the 12-hour format.
COMMUNICATION RESET
Changes the communication settings to the default factory values listed in
Table 15 on page 202.
Table 14 Measurement default settings
Parameter
Default value
Measurement mode with differential
Fast, min, max, and alarms turned off
Material sound velocity
5.969 mm/s (or 0.2350 in./s)
(approximate velocity of the included test
blocks)
Differential reference value
0.0 mm (or 0.0 in.)
Low alarm reference value
0.0 mm (or 0.0 in.)
High alarm reference value
635.0 mm (or 25.0 in.)
Display update rate
4 per second
Zoom Off
Range
Minimum range
LOS condition
Blank thickness display
Resolution
0.01 mm (or 0.001 in.)
CAUTION
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...