DMTA-10022-01EN, Rev. C, January 2015
Chapter 7
128
The units and resolution of the differential thickness are the same as those selected for
the thickness measurement.
When you press
[SAVE]
(with the optional datalogger) while in the
NORMAL
or
% RATIO
differential modes, the 45MG saves the actual thickness value along with
the “D” flag indicating that the
Differential
mode is active.
To activate and configure a differential mode
1.
In the measurement screen, press
[SETUP]
, and then select
DIFF
.
2.
In the
DIFF
screen (see Figure 7-2 on page 129):
a)
Set
DIFF ENABLE
to
ON
to activate the differential function.
b)
In
DIFF MODE
, select one of the three differential modes:
•
NORMAL
: Shows the actual thickness, along with the difference between
the actual thickness measurement and the
REF VALUE
entered.
•
% RATIO
: Shows the actual thickness along with the percent difference
from the
REF VALUE
entered.
•
REDUCTION RT
: Shows the actual thickness, in addition to the percent
difference between the actual thickness and the former value. The former
value is the thickness of the metal prior to the bending process. Use this
mode for metal bending, or any other application in which you need to
track the percentage of wall thinning.
c)
When
DIFF MODE
is set to either
NORMAL
or
% RATIO
, enter the
reference value in the
REF VALUE
text box.
Differential
Normal
Current thickness Reference value
–
=
Differential
% Ratio
Current thickness Reference value
–
Reference value
--------------------------------------------------------------------------------------- 100
×
=
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...