DMTA-10004-01EN, Rev. D, November 2016
Using Advanced Gaging Features
157
d)
Set
TEMP COEFFICIENT
to the desired value (typically –0.00018 for °C or –
0.00010 for °F) for steel.
e)
Set
CURRENT TEMP
to the temperature of the actual test piece.
You can specify a negative sign using the
[
]
or
[
]
keys to move the cursor over the
“+” or “–”, and then using the
[
]
and
[
]
keys to change the sign.
Figure 10
‑
18 The TEMP COMP screen
4.
Press
[MEAS]
to return to the measurement screen with the activated
temperature compensation (see Figure 10-19 on page 158).
The measured thickness is temperature-compensated based on the
CALIBRATION TEMP
, the
CURRENT TEMP
, and the
TEMP COEFFICIENT
parameter values.
NOTE
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...