Figure 8
‑
5 Example of a B
‑
scan alarm mode
The thickness value and the alarm indicator appear in color only when the indoor
color scheme is active (see “Color Schemes” on page 52 to change the color scheme).
REDUCTION RT
The
REDUCTION RT
option appears only when the active file is configured with
FILE DATA MODE
set to
REDUCTION RT
. You set the reduction percentage for
a low (
YELLOW ALARM
) and high (
RED ALARM
) limits, and the instrument
displays (see Figure 8-6 on page 122):
— The
RED
indicator for reduction rates that are greater than or equal to the
RED ALARM
value.
— The
YEL
indicator for reduction rates that are between the
YELLOW ALARM
and the
RED ALARM
values.
— The
GRN
indicator for reduction rates that are below the
YELLOW ALARM
value.
Red alarm limits
NOTE
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...