DMTA-10004-01EN, Rev. D, November 2016
Using EMAT Transducers
89
6.2
Calibrating with the E110-SB EMAT Transducer
Calibration is the adjustment of the gage so that it measures accurately on a particular
material. The default velocity and zero offset for the E110-SB EMAT transducer are
designed to provide a good estimate of the ferrous metal thickness underneath an
external oxide scale coating.
To achieve the best accuracy, use calibration samples with external oxide scale made
from the material to be tested. These samples should have a metal of known thickness
(not including external oxide scale) and should represent the minimum and
maximum thickness of your measurement range.
The calibration procedure is the same as for a standard transducer except that you do
not need to use couplant to couple the transducer to the material. Refer to
“Calibrating the Instrument” on page 67 for the calibration procedure.
The EMAT transducer generates shear wave sound energy. Consequently, the velocity
calculated by the gage is the shear sound velocity of the material.
NOTE
Summary of Contents for 38DL PLUS
Page 9: ...DMTA 10004 01EN Rev D November 2016 Table of Contents ix Index 297 ...
Page 10: ...DMTA 10004 01EN Rev D November 2016 Table of Contents x ...
Page 28: ...DMTA 10004 01EN Rev D November 2016 Important Information Please Read Before Use 16 ...
Page 30: ...DMTA 10004 01EN Rev D November 2016 Introduction 18 ...
Page 50: ...DMTA 10004 01EN Rev D November 2016 Chapter 2 38 ...
Page 60: ...DMTA 10004 01EN Rev D November 2016 Chapter 3 48 ...
Page 102: ...DMTA 10004 01EN Rev D November 2016 Chapter 6 90 ...
Page 222: ...DMTA 10004 01EN Rev D November 2016 Chapter 11 210 ...
Page 254: ...DMTA 10004 01EN Rev D November 2016 Chapter 13 242 ...
Page 274: ...DMTA 10004 01EN Rev D November 2016 Chapter 14 262 ...
Page 286: ...DMTA 10004 01EN Rev D November 2016 Chapter 15 274 ...
Page 294: ...DMTA 10004 01EN Rev D November 2016 Appendix A 282 ...
Page 300: ...DMTA 10004 01EN Rev D November 2016 Appendix C 288 ...
Page 306: ...DMTA 10004 01EN Rev D November 2016 List of Figures 294 ...