Olympic Medical OLYMPIC SMART SCALE 35 Operating And Service Instructions Download Page 1

Summary of Contents for OLYMPIC SMART SCALE 35

Page 1: ...OLYMPIC SMART SCALRtm Model 35 Operating and Service Instructions...

Page 2: ...erating and Service Instructions Table of Contents Page Summary Operating Instructions 1 1 Safety Precautions 2 1 Set Up 3 1 Operating Instructions 4 1 Service 5 1 OLYMPIC MEDICAL 4400 Seventh South S...

Page 3: ...e or placing baby on scale To weigh baby Press Display button to turn display on Press Gms Lbs button to select units Press Lock button to lock average weight computed during last 4 seconds on display...

Page 4: ...weighing platform in place 4 The scaleisdesigned foruseonlywiththe babyconfined in an incubator or radiantwarmer Do not use weighing platform on an open tabletop or for carrying the baby Never permit...

Page 5: ...ny other warmer with opening in bed base without using this adapter CAUTION When using scale with opening in bed base adapter base must be used For radiant warmer use assemble weighing platform as sho...

Page 6: ...ays by sliding them all the way in the long slot then along short slot as shown in Fig 3 4 Lift off Clean with liquid disinfectant or mild soap and water detergent Re install trays by setting rounded...

Page 7: ...back of Display Control Unit When used in incubator pass cable through access hole Connect power cord to properly grounded hospital grade electrical outlet 115 Vac 60 Hz If the Display Control Unit i...

Page 8: ...ro button also causes scale to find its zero reference point to assure accurate measurements When Zero button is pressed moving dashes will appear on the display for a few moments Wait until the dashe...

Page 9: ...d average Then lay tubes and leads down and chart weight 2 Use Lock Average Button While you are holding the tubas and leads have another nurse push the LocknAverage button The best average weight com...

Page 10: ...ightly Let them rest in the open palms of your hands while lifting 4 Gently raise your hands to the point where the tubes and leads are just lifted off the baby s body but no higher Lifting higher tha...

Page 11: ......

Page 12: ...Exercise the weighing platform load cell by placing the 5 kg calibration weight in the center for 10 seconds then remove the weight for 10 seconds Repeat 10 times 5 Connect the power cord to electric...

Page 13: ...RO Display should read 12 768 4 If not repeat steps 7 thru 13 omitting the coarse adjustments of steps 8 and 12 m m 15 Put all DIP switches in the Off down position and replace the calibration cover 1...

Page 14: ...to signal ground TP4 on the Digital board Static and Contamination Awareness Most integrated circuits and semiconductors in this Scale can be damaged by electrostatic discharge Exercise proper handli...

Page 15: ...ed screws in the rear 4 Reassemble clamp adjust spacing 80 away from enclosure I 7 Tighten the central screw 8 Replace the calibration cover Removing Analog Digital PCBs w ViH fi Make sure electric po...

Page 16: ...bad replace Digital board Scale will not lock the display and the READ light will not come on 1 Check that the DIP switch Si 4 on Analog board is off down 2 Pressing LOCK AVG should provide a low to...

Page 17: ...onverter is 134 microvolts per gram I I I The converter internally resolves to Va gram It divides 10 kilograms maximum weight for scale into 40 000 counts 20 000 range A D gain is controiied by R18 an...

Page 18: ...tween appropriate terminals of El it is possible to measure current of a group of lit segments Power Supply Power supply for the Smart Scale Model 35 is divided between the two major printed circuit b...

Page 19: ......

Page 20: ......

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...TP1 TP2 TP4...

Page 25: ...TRANSFORMe R f LTeR A55y...

Page 26: ......

Page 27: ......