46
Set Point 37.0°C
Base
Err_2
HM 37.0
Lid
37.0
OBH 37.0
Figure 36. Temperature sensor fault.
In Figure 39 above, ERR_2 means that there is no temperature reading form the base of the
chamber. This happens if the sensor embedded into the base plate of the chamber is faulty or
if its cable is disconnected from the outlet labeled T1. The same ERR_2 appears if Lid or
Humidity Module sensor is faulty or disconnected.
Screen Settings Submenu
Menu 2/3
Humidity Module
Alarm Settings
à
Screen
Settings
8
Screen Settings
à
Contrast: 50
Brightness: 08
Reset to factory
Change
the value
-
Exit
8
Confirm
new value
â
á
Select the value
to be changed
â
á
Move the cursor to the
Parameter to be
changed
Figure 37. Screen Settings Menu
The Screen Settings Sub Menu enables the operator to adjust the instrument screen contrast
and brightness. Use the option ‘Reset to factory’ to set Screen Settings back to the default
values.
From the main menu press the Menu key on the front panel.
Move the cursor to ‘Screen Settings’ using the Arrow Keys.
Press ‘Enter’.